Shimadzu GC-2014C
4 Using the control module
7
4.2 Method Setup - GC
4.2.1 Oven/Zones
Menu Path:
Instrument -> Method -> GC Control
Fig. 7 Method Setup – GC – Oven/Zones
Oven Max
Sets the maximal allowed temperature for the column oven.
Possible values are:
0
-
440
.
Please note that the maximum value may differ for certain GC
models - listed values for it range between
400
and
470
°C.
Equilibry Time
Time that the GC needs to reach the temperature
equilibrium after the temperature program gets to the set
initial temperature. This time is used to distribute the heat
evenly in the oven.
Gradient Table
Sets the
Heat Rate
, desired temperature (
Final Temp
) and
time to hold this temperature (
Hold Time
). The
Total Time
column will be calculated automatically. In other words, for
isothermal part of the temperature gradient, the temperature
set in
Final Temp
will be kept for the interval set in
Hold