INT7 A/D Converter
5 Troubleshooting
5.4 Data Acquisition - non-functional
▌ Gray
icon with the heading
DISABLED
and non- functional
Monitor - Data
Acquisition
command.
Fig 33: Data Acquisition disabled
Other manifestations of this error are also:
tab missing,
Method - Acquisition
command non-functional,
Run
,
Stop
,
Abort
and other similar
commands non- functional in the
,
or
windows. Possible causes are:
Description:
a) You are using
Clarity Offline
or a
Clarity DEMO
version, which does not
enable the measurement of chromatograms.
Fig 34: Clarity Offline and Clarity DEMO
Solution:
Check whether there is a blue line with the title
OFFLINE
displayed in the main
window under the symbols of the Instruments, or the title
DEMO
in the
window header.
In case of
Clarity DEMO
, use the
Windows Start menu
to locate the
Clarity
DEMO
group in the
All Programs
section and select the
Remove Clarity
item there.
Then, use the correct
Clarity
installation DVD to install the full version.
In case of the
Clarity Offline
station, remove the HW key with the
Offline
license
and insert key with the
Clarity
full licence.
Description:
b) Detector is not allocated to the
Clarity
Instrument:
Fig 35: Detector not allocated to the Instrument
Solution:
Open the
dialog from the
window using the
System -
Configuration...
command and check the tab of the corresponding Instrument -
Instrument X
. If it has no allocated detectors, add them.
In the left-hand list
Setup Control Modules
select the correct detector connected to
the
A/D card
you are using and drag it to the corresponding instrument on the
right.
- 32 -