Calibration
85
Calibrating the Analog Output Subsystem
Once the DT9850 Series Calibration Utility is running and you have connected a precision
digital multimeter (DMM) to the module’s outputs, do the following to calibrate the analog
output subsystem:
1.
If you have not already done so, select the name of the module to calibrate from the
Board
drop-down list box. The utility lists only DT9850 Series modules.
2.
Select one of the
Output Range
buttons:
-10V to 10V
(bipolar) or
0V to 10V
(unipolar).
Note:
If you are solely interested in using current output, calibrate the unipolar voltage
range.
3.
Select
DA channel 0
. (The list box reflects the number of channels available on the
module.)
4.
Select the
Offset calibration
button.
5.
Click the increment or decrement arrows in the adjustment box until the measured
voltage on the connected DMM is 0 V (within 0.0005 V).
6.
Select the
Gain Calibration
button.
7.
Click the increment or decrement arrows in the adjustment box until the measured
voltage on the connected DMM is 0 V (within 0.0005 V).
8.
Select the next
DA channel
to calibrate.
9.
Repeat steps 4 to 7 until all the DACs on the module are calibrated.
10.
Click
Done
when you are finished calibrating the analog output circuitry.
Summary of Contents for DT9850 Series
Page 1: ...DT9850 Series UM 22600 M User s Manual Title Page ...
Page 4: ......
Page 8: ...Contents 8 ...
Page 16: ...Chapter 1 16 ...
Page 17: ...Part 1 Getting Started ...
Page 18: ......
Page 48: ...Chapter 4 48 ...
Page 49: ...Part 2 Using Your Module ...
Page 50: ......
Page 76: ...Chapter 6 76 ...
Page 82: ...Chapter 7 82 ...
Page 83: ...83 8 Calibration Using the Calibration Utility 84 Calibrating the Analog Output Subsystem 85 ...
Page 86: ...Chapter 8 86 ...