Chapter 1
24
Figure 3: IEPE Block Diagram on the DT9841-VIB
Input Resolution
The input resolution of the DT9841, DT9841E, and DT9841-VIB modules is fixed at 24-bits. The
input resolution of the DT9842/2 and DT9842/8 modules is fixed at 16-bits.
Input Range and Gain
All DT9840 Series modules provide an input range of ±10 V and a fixed gain of 1.
Data Format
The analog input data encoding used by the DT9841, DT9841E, and DT9841-VIB modules is
floating point. The data is 32-bit wide and left justified; the right eight bits are filled with
zeros.
The DT9842/2 and DT9842/8 modules use twos complement data encoding. The data is 32-bit
wide and left justified; the right 16 bits are filled with zeros.
Note:
When the module is above range, the value FFFFh (plus full-scale) is returned. When
the module is below range, the value 0000h (minus full-scale) is returned.
You can use software to convert a twos complement code into voltage.
DC-DC
Coupling
AC - DC
4mA
BNC 0
x1
2 Pole
Butterworth
Filter
10kHz
+30V
Filter
ON - OFF
DT9841-ICP Block Diagram
+5V
Ch0
1.6Hz
DIO Port 2
Bit 0
1 of 8
Channels
1 of 8
Channels
DIO Port 2
Bit 1
SDA
SCL
MAX7300
Channel 1
Channel 2
Channel 3
Channel 4
Channel 5
Channel 6
Channel 7
P30
P29
P31
P28,27,26
P25,24,23
P22,21,20
P19,18,17
P16,15,14
P13,12,11
P10,9,8
Control
Lines