Setup ■ Machine Installation
- 34 -
Data I/O ■ 096-0465-001C
the floor, or if the leg locknuts are not tightened against the frame. Only the
adjustable feet provided on the PSV5000 Machine should be used for
leveling. Do not use shims to assist in leveling.
Connecting Facilities
The PSV5000 requires two external services: pressurized air and
electrical power. It also allows one optional connection to an Internet via
a Network connection. These services connect to the PSV5000 Machine
at the Power Panel located on the back of the machine.
With the two required services, the machine creates all the unique
electrical voltages needed as well regulating pneumatic pressure for all
systems within the machine.
Connecting Air
The machine requires clean, dry, oil-free air from an industrial grade
compressor.
PSV5000 Air Requirements Table
Note:
If the compressor cannot maintain the correct air pressure and
stable air volume, PSV5000 System performance will be affected and may
damage devices.
System
Air Volume
@ Air
Pressure
at inlet
regulator
on
PSV5000
Supply Air
Min.
Air
Tube
Length
1
Customer supplies a
primary
filter/regulator
required between
factory compressor
and PSV50002
—Metric
85 L/minute
[.084 m3/min],
t t
5.5 ±0.3 Bar 5.5
‐
8.2 Bar 3 Meters
10 micron
3 SCFM constant 80 ±5 PSI
80 to 120 PSI 10 Feet
10 micron
1
This external air line must allow the supplied air to cool sufficiently so that water vapor in
the air condenses and can be extracted.
2
The air filter on the Facilities Panel is a secondary filter only.
Summary of Contents for PSV5000
Page 12: ......
Page 36: ......
Page 160: ......
Page 161: ...Electrical Schematics for PSV5000 and Equipment ...
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ...PROGRAMMABLE MEDIA EXPERTS ...