background image

- 5 -

3.5 Pipeline mounting

Connect the gray backplane to the instrument, and then fix the backplane to the horizontal

pipe or vertical pipe with a clamp. The installation diagram is as follows:

Summary of Contents for MCC100

Page 1: ...Multi parameter Controller Model MCC100 200 Version 1 0 User Manual Daruifuno...

Page 2: ......

Page 3: ...n with particular attention to all danger warnings and precautions Failure to do so may result in damage to the instrument or personal injury to the operator due to misuse Do not install or use the in...

Page 4: ...er this warranty is limited to providing replacement or repair of this product as the case may be The product must be thoroughly cleaned to remove any contaminated chemicals before it is accepted for...

Page 5: ...tton 8 5 2 Display 9 Chapter 7 Menu Description 4 1 Power connection 6 4 2 Wiring terminals definition 7 Chapter 4 Electrical Installation 1 1 Technical parameters 1 3 1 Installation location 2 3 2 Di...

Page 6: ...4 4 Lower value 13 7 5 Communication set 13 7 5 1 Slave ID 13 7 5 2 Baud rate 13 7 5 3 Verify bit 13 7 5 4 Stop bit 13 7 6 System set 13 7 6 1 Language 13 7 6 2 Display speed 13 7 6 3 Backlight 13 7 6...

Page 7: ...ambient temperature alarm Contact type SPST Contact capacity 3A 250VAC Current output Number of channels 2 Output type active 4 20mA or 0 20mA Maximum load 1000 Communication interface RS485 MODBUS R...

Page 8: ...power spike interference if power spike interference occurs the power supply of the controller can be separated from the power supply of the control device so that the controller uses a separate powe...

Page 9: ...panel cabinet type use two fixing brackets to fix the instrument The installation diagram is as follows 3 3 Panel mounting Figure 3 2 Panel mounting 3 2 Dimension Figure 3 1 Schematic diagram of the a...

Page 10: ...the instrument and then install it on the box or the side wall of the wall The installation dimensions and schematic diagram are as follows Figure 3 3 Schematic diagram of wall mounted installation Fi...

Page 11: ...5 3 5 Pipeline mounting Connect the gray backplane to the instrument and then fix the backplane to the horizontal pipe or vertical pipe with a clamp The installation diagram is as follows...

Page 12: ...n see two rows of terminals Figure 4 1 indicates the location and name of the access terminals for both power supplies AC power supply 100 240VAC Two kinds of power terminals are defined in Table 4 1...

Page 13: ...V Electrode power terminal Current two output negative V Electrode power terminal OUT1 Current one output positive BS Electrode communication terminal B Current one output negative R1 Relay one conta...

Page 14: ...the value or to change the sign bit In measurement mode press this key to switch between two measurement value display modes only pH measurement mode is available In the menu press this key to move t...

Page 15: ...Ambient temperature alarm When the operating environment temperature exceeds 50 a will appear on the screen indicating that the working environment temperature of the instrument is too high and heat d...

Page 16: ...I Output Type 4 20mA Current II Output Type 4 20mA Comm Set Net Address 001 Baud Rate 9600 Verify Bit NONE Stop Bit 1 bit System Set Display Speed Buffers 8 Backlight Scanty Power 01 Minute Contrast S...

Page 17: ...ter the target temperature value and press the key to calibrate the temperature measurement to the entered target value Reset the temperature calibration data to the factory default values 7 2 3 Tempe...

Page 18: ...gs Relay X Function Settings menu the relay is in the alarm control output state The relay status is controlled by the working environment temperature of the instrument When the environment temperatur...

Page 19: ...the menu in the measurement mode The password is set to 000000 when leaving the factory and you can enter the menu directly without password in the measurement mode If the user sets other password the...

Page 20: ...zzer alarm Set whether the buzzer outputs an alarm sound when the ambient temperature is alarmed Two current outputs can be tested First connect the ammeter to the current output terminal to be tested...

Page 21: ...setting Set a value in the current test menu and check whether the controller output is normal Please contact technical support Inaccurate current output Check current output setting Recalibrate the...

Page 22: ...ssion process the entire message frame must be transmitted in a continuous stream If there is a pause interval of more than 1 5 characters before the completion of the message frame transmission the r...

Page 23: ...00 00 The starting register address is 0x0000 00 02 Read 2 registers starting from the starting address that is read 1 single precision floating point data result 71 CB CRC check data Return frame 01...

Page 24: ...Suzhou Delfino Environmental Technology CO Ltd www daruifuno com info daruifuno com...

Reviews: