4.1.1 Considerations
Wall Construction
Before fixing it is important to consider the strength of
the wall and the depth of the fixings used.
Use of a building engineer or structural drawings may be
required to decide on suitability.
Examinations to be
undertaken
We recommend the bracket be installed approximately
850-1100mm from the floor to the underside of the
bracket – dependent on couch height.
It may be necessary to adjust this height, if there are
specific medical procedures the light will be used for. If
in doubt consult the end user or DARAY. When installing
the bracket take consideration of where you want the
light head to sit when in its lowest position.
Locations of nearest
mains sockets
The X740 wall mounted light uses a plug-top transformer
with a DC jack connection.
Only the supplied certified power adapter should be used
with the light.
The length of this cable is approximately 1.8m so the wall
mounted light should be within 1.5m of a mains power
point.
NOTE: The following versions are not supplied with a
plug-top transformer (X740LE1 / X740LE2 / X740LE3 /
X740LE4). A separate transformer is supplied to make a
hardwired connection.