6. Troubleshooting Guide & Spare Parts 

 

Problem 

Possible Cause 

Corrective Actions 

 

No light output 

Power Supply 

Wall Version - Check the plug top 
adapter is plugged in and switched 
on at the power socket. 

On/Off/Intensity 

Button 

Use the control panel to check if the 
light will power on (if dimmed refer 
to poor light output actions). 

Fuses 

Visually check fuses then check 
continuity with meter 

Cables 

Check the visible wiring and make 
sure the DC connectors between the 
light and the extension cable are 
fully connected. For the mobile 
version, ensure that the IEC is firmly 
pushed into the PSU. 

 

Poor Light 

output 

Power Button 

Please check the variable intensity is 
not just set to its minimum. To do 
this use the intensity controls on the 
side of the light head. 

Voltage in 

Verify that the correct mains voltage 
and correct transformer is being 
used  

Voltage out 

Verify correct secondary voltage 

Arm drift 

sideways 

Wall mount 

On installation the arm has most 
likely not been vertically aligned and 
levelled. Please contact the original 
fitter or a qualified electrician 

 
 
 
 
 
 

 

Summary of Contents for X700

Page 1: ...X700 LED Examination Light Operating Installation Manual QAM X700 0519 4...

Page 2: ...Page 1 of 24...

Page 3: ...4 1 1 Considerations 10 4 1 2 Placement and Installation 11 4 2 Rail mounted version X700LR 13 4 3 Desk mounted versions X700LD X700LDS X700LFDM 13 4 4 Mobile version X700LM 14 4 5 Optional battery b...

Page 4: ...e light patch The light comprises 3 parts light head and flexible arm system power supply module and the mount of choice The flexible arm allows the light patch to be placed accurately and stably wher...

Page 5: ...ured in Great Britain by Daray Ltd Edison House Robian Way Swadlincote Derbyshire DE11 9DH 1 1 Supported Lights The Following Models are supported in this Manual X700LW X700 LED Wall Mounted Light X70...

Page 6: ...000K 500 Colour Rendering Index CRI 95 Colour Rendering R9 95 Light Field Diameter D50 0 5m 60mm 0 75m 90mm 1 0m 120mm Light Field Diameter D10 0 5m 110mm 0 75m 150mm 1 0m 200mm Number of LED s 4 Powe...

Page 7: ...top transformer 1 8m 1 Desk Clamp Fixed Desk Mount Desk Stand 1 Wall bracket fixings 1 Light head and flexible arm 1 Thumb Screw for Wall Mount 1 Microfibre cloth 1 Light head and flexible arm 1 User...

Page 8: ...Page 7 of 24 3 Operation 3 1 Range of Motion This is an image depicting the range of motion on the X700 light all versions...

Page 9: ...Page 8 of 24...

Page 10: ...mperature Press the power button to power ON or OFF Press and hold the plus or minus buttons to adjust the intensity It takes approximately 5 seconds to switch between the highest and lowest settings...

Page 11: ...r to the underside of the bracket dependent on couch height It may be necessary to adjust this height if there are specific medical procedures the light will be used for If in doubt consult the end us...

Page 12: ...in mind the final position of the light head required to project its light patch where it will be needed We recommend 850 1100mm from the floor to the underside of the bracket dependent on couch heig...

Page 13: ...light on See page 9 Check arm and head for smooth and correct movement Your new DARAY light is now ready for use To gain full support of our warranty please fill in our warranty web form or contact DA...

Page 14: ...can be moved along the rail by slackening the clamp slightly and not sufficiently to allow the light to become detached from the rail 4 3 Desk mounted versions X700LD X700LDS X700LFDM X700LD includes...

Page 15: ...g If not already fitted fit the castors to the base by knocking them sharply into their fixing sockets Insert the up stand into the castor base as shown below Do not use a metal hammer Once fitted the...

Page 16: ...e into the black power supply attached to the base upstand The light can now be tested by connecting the DC jacks between the power supply and the light and applying power Press the switch on the head...

Page 17: ...mobile version of this light please use the extension cable provided NOTE Charging time can take up to 8 hours depending on how depleted the battery is 2 If you need to use the light on battery power...

Page 18: ...ge Remove arm clean spigot and re grease as necessary Check all electrical connections are sound and that there is no visible cable wear All fixings to be checked and be of sound construction 5 2 Safe...

Page 19: ...ating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment and accessories included charger...

Page 20: ...ight 5 4 2 Detaching Parts De construction of the lights should be done following a reverse order of the instructions in section 4 The X700 is a completely enclosed light and arm it should not be open...

Page 21: ...make sure the DC connectors between the light and the extension cable are fully connected For the mobile version ensure that the IEC is firmly pushed into the PSU Poor Light output Power Button Please...

Page 22: ...SPARE PARTS or TECHNICAL ISSUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am 5pm Monday to Thursday excluding bank holidays 9am 2 30pm on Friday Or email support daray c...

Page 23: ...ges may be applicable OTHER Any situation which is not covered by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our ter...

Page 24: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Page 25: ...Page 24 of 24 Service Notes...