Maintenance & Storage
Cleaning
•
Use a soft cloth and medical alcohol to clean the
device, and allow to air dry.
Maintenance
•Cleanthedeviceregularlyasperabove
•Keepthebatterychargedandrechargewhen
prompted by the V406
•Conditionthebatteryevery6months,especially
during periods when the device is rarely used.
Storage
•TheV406isbeststoredinoriginalpackaging
•
Remove the battery during long storage periods
•Donotstorewithcorrosiveorexplosivegases
•Storewithplentyofventilation
•Storagetemperaturerange:-40to60°C
Quick Setup Guide
Problem
Cause
Solution
Abnormal or eratic
SpO2/PRreadings
Thepatientsfingerisnotcorrectlypositioned
Removethepatientsfingerandreinsert
ThepatientsSpO2/PRistoolowtobedetected
Removepatientsfingerandattemptreadingagain.
Thepatientismoving/shaking
askpatienttoremainasstillaspossible
The patient is wearing nail varnish
Removeanynailpolishfromthepatientsfinger
V406 is not
powering on
The battery has little or no charge
Charge the battery for at least 8 hours using the
charger provided
The battery is not installed or is not correctly
installed
Checkthebatteryisinstalledandthatthepolarity
is correct
The V406 is faulty
Return to DARAY for inspection and repair
The display
powers off
suddenly
The battery has little or no charge
Charge the battery for at least 8 hours using the
charger provided
The V406 is set to power off after 5 seconds when
not in use
Normal operation
The unit can not
be used for long
periods of time
The battery has little or no charge
Charge the battery for at least 8 hours using the
charger provided
The battery is damaged
Contact DARAY for replacement
Troubleshooting
Specification
Sp02
Range
0-100%
(1%Resolution)
Accuracy
±2%at70-100%
PulseRate
Range
30-250BPM
(1BPMResolution)
Accuracy
±2BPMor±2%
Dimensions
46(L)x40(W)x29(H)mm
Weight
Approx.35g(inc.battery)
Operating Environment
Temp.
10°C-40°C
Humidity
30%-75%
Pressure
700hPa1060hPa
InfaredWorkingElements