DARAY V406 User Manual Download Page 2

DARAY Ltd. Marquis Drive, Moira, Derbyshire DE12 6EJ

Tel:  +44 (0)333 321 0971     Fax: +44 (0)333 321 0973

[email protected]       www.daray.com

V406 User Manual

DOCUMENTNo.QAM.SUG.V406.1012.1

DARAY Ltd. registered in England no. 05276678

Maintenance & Storage

Cleaning 

Use a soft cloth and medical alcohol to clean the 

device, and allow to air dry.

Maintenance

•Cleanthedeviceregularlyasperabove
•Keepthebatterychargedandrechargewhen

prompted by the V406

•Conditionthebatteryevery6months,especially

 

   during periods when the device is rarely used.

Storage

•TheV406isbeststoredinoriginalpackaging

Remove the battery during long storage periods

•Donotstorewithcorrosiveorexplosivegases
•Storewithplentyofventilation
•Storagetemperaturerange:-40to60°C

Quick Setup Guide

Problem

Cause

Solution

Abnormal or eratic 

SpO2/PRreadings

Thepatientsfingerisnotcorrectlypositioned

Removethepatientsfingerandreinsert

ThepatientsSpO2/PRistoolowtobedetected

Removepatientsfingerandattemptreadingagain.

Thepatientismoving/shaking

askpatienttoremainasstillaspossible

The patient is wearing nail varnish

Removeanynailpolishfromthepatientsfinger

V406 is not 

powering on

The battery has little or no charge

Charge the battery for at least 8 hours using the 

charger provided

The battery is not installed or is not correctly 

installed

Checkthebatteryisinstalledandthatthepolarity

is correct

The V406 is faulty

Return to DARAY for inspection and repair

The display 

powers off 

suddenly

The battery has little or no charge

Charge the battery for at least 8 hours using the 

charger provided

The V406 is set to power off after 5 seconds when 

not in use

Normal operation

The unit can not 

be used for long 

periods of time

The battery has little or no charge

Charge the battery for at least 8 hours using the 

charger provided

The battery is damaged

Contact DARAY for replacement

Troubleshooting

Specification

Sp02

Range

0-100%
(1%Resolution)

Accuracy

±2%at70-100%

PulseRate

Range

30-250BPM
(1BPMResolution)

Accuracy

±2BPMor±2%

Dimensions

46(L)x40(W)x29(H)mm

Weight

Approx.35g(inc.battery)

Operating Environment
Temp.

10°C-40°C

Humidity

30%-75%

Pressure

700hPa1060hPa

InfaredWorkingElements