APPLICATION INFORMATION
C
onnect the test set to other systems using the 05615 Univer-
sal Alarm Test Set Cables Package. Refer to Fig. 11. The
test set has multiple ports for communications with various
external alarm systems. Refer to Fig. 12. Use the types of cables
for connecting to different systems.
Fig. 13 shows a standalone alarm interrogator application. The
test set automatically interrogates alarms based on
preprogrammed instructions stored in non-volatile RAM
memory. The processed alarms route to the master computer
upon request. The MAP printer port has alarm information in
ASCII format.
Fig. 14 shows a TABS/TBOS responder application. The test set
connects to an alarm master system using TABS, TBOS or
another protocol through port C2 of the 46090 RS-232 Connec-
tor. Port C1 of the 46090 RS-232 Connector connects to a local
configuration or monitor port.
Use the 46022-3X Multiple Alarm Combiner for a four wire
connection to a local TABS or TBOS device being tested. You can
also use the 44210 460 Modem for a dial-up configuration.
Fig. 15 shows a TABS/TBOS interrogator application. The test
set connects to an alarm master system using DCP, DCPF or
another protocol through port C2 of the 46090 RS-232 Connec-
tor. All other connections are the same as the TABS/TBOS
responder application.
The test set automatically interrogates alarms based on
preprogrammed instructions stored in non-volatile RAM
memory. The processed alarms route to the master computer
upon request. The MAP printer port has alarm information in
ASCII format.