DANA Diabecare DANA-i Professional Manual Download Page 1

Professional Manual

Summary of Contents for Diabecare DANA-i

Page 1: ...Professional Manual...

Page 2: ...tor Mode 3 What can be adjusted within Dcotor Mode 4 3 Password Mode 9 Accessing Password Mode 9 What can be adjusted within Password Mode 10 4 Distributor Mode 11 Accessing Distributor Mode 11 What c...

Page 3: ...e insulin pump configuration The information within this manual is supplementary to the IFU that is distributed with the DANA Insulin Pump Warning This is not for distribution to patients or end users...

Page 4: ...ld three buttons at the same time 2 The pump will request a PASSWORD enter 30XY and press OK Notice THE XY of 30XY is derived from the manufacturing day of the pumps You can see the manufacturing date...

Page 5: ...rbohydrate meals for breakfast lunch and dinner Set and save 3 different insulin dosages for meal bolus of breakfast lunch and dinner On pre set bolus is activated Off pre set bolus is deactivated Pre...

Page 6: ...n Bolus block is activated symbol will be shown on the initial screen Bolus Block can be set between 0 180 minutes in 10 minute intervals To deactivate the bolus block feature set the time to 0 zero B...

Page 7: ...an hourly absorption ratio for insulin For the DANA insulin pump to calculate for residual active insulin the DEC RATIO to suit the individual patient needs to be set The DEC RATIO can be set from 5 5...

Page 8: ...ILY MAXIMUM can be set between 0u to 300u in 1u increments Safety Ratio This is a safety feature to prevent over dose of insulin from wireless commands BT enabled devices When the bolus is administrat...

Page 9: ...nd is usually adjusted for total daily dose UTC SETTING Set default UTC Time Date and time used in the OPTION CLOCK SETTING EXIT Save all setting and exit the menu It is necessary to exit from the Dr...

Page 10: ...N MENU press and hold the three buttons at the same time 2 The pump will request a PASSWORD enter 3571 as the following sequence 1 3 5 7 and press OK Notice Must input the last number 1 first if not c...

Page 11: ...apter 2 1 CLEAR US 4020 This is a password to delete the history of the day total dose Useful to reset or clear the already delivered daily dose input 4020 on the password screen after press the three...

Page 12: ...IN MENU press and hold the three buttons at the same time 2 The pump will request a PASSWORD enter 8949 as the following sequence 4 9 8 9 and press OK Notice Must input the third number 4 first or can...

Page 13: ...here BLOCK SENSITIVE Is a variable sensitivity for any occlusion within the pump or infusion sets Occlusions occur due to crystallization blood or infusion set kink age The Block sensitivity can be a...

Page 14: ...G MODE of the DANA Insulin pump it is therefore possible to adjust the bolus audio tone from melody to beep or silence Warning This is not for distribution to patients end users or diabetes educators...

Page 15: ...14 MEMO...

Page 16: ...15 IFP 130 EN rev 0_200701...