EMC compliance
To ensure EMC compliance, follow these specific guidelines:
Ensure that all cable shields have a 360
electrical connection to the connector.
Fasten and secure all connectors.
Warning
The detector can influence the performance of very sensitive equipment.
Strong EMC disturbances can influence the detector’s performance.
3.
System precautions
The following are precautions that must be taken in order to prevent possible damage to your Teledyne DALSA
Detector system:
Warning:
The use of accessories, power supplies and cables other than those supplied or specified, with the exception of
cables sold by Teledyne DALSA as replacement parts, may result in increased emission or decreased immunity of
the Shad-o-Box HS Detector.
Static precautions
Observe proper ESD/static control procedures when handling system
components. The use of properly grounded wrist strap is highly recommended.
Installation Precautions
Never connect or disconnect cables while power to the detector is on. Damage to the detector electronics will occur
if the cables are connected and/or disconnected while the detectors power is on.
Service by Qualified Personnel Only
This equipment is to be installed by a qualified technician only. The system can only be used in conjunction with a
properly installed X-ray source with the appropriate shielding and a properly configured computer workstation that
meets the minimum system requirements discussed later in this manual.