Renewing the Validity of the Calibration.
Usually, the validity of the calibration settings for a detector are renewed automatically when
it is calibrated by a technician certified by
DALEMANS
. The maintenance interval for the
calibrated detector is then reset.
In some specific cases, a detector may need to be verified by a certified control authority and
the validity of the calibration of this detector may then be extended.
The user can select the detector for which the validity of the calibration is to be renewed
using the
Up
and
Down
arrow keys. Each detector is displayed with the number of days that
have lapsed since its last calibration and the full-scale value of the gas concentration that the
detector is able to estimate.
1.Detector 1
346 days 98%LEL
The user can confirm their selection of detector by pressing
ENTER
. The system then asks
the user to confirm the renewal of the validity of the calibration for the selected detector.
1.Detector 1
Renew calibration ?
The user can confirm the renewal of the validity of the calibration for the detector by
pressing
ENTER
. The next maintenance date may have been modified at this point. The
user can cancel the modification and return to the info menu by pressing
ESC
.
Detecting the Detectors
Navigation :
Service menu
→
Configuration menu
→
Sensor detection
The factory settings of the U•V6 unit disable all the channels. The U•V6 unit then needs to
be told how many detectors are actually connected.
This sub-
menu can be used to get the U•V6 unit to verify the channels and display the
situation as it sees it.
Detection and Calibration Phase
▪
A rotating symbol shows the progress of the calibration.
▪
A "-" (hyphen) symbol indicates a channel that does not have a detector or a channel
that is connected to a detector for which the zero calibration failed (too much signal
interference);
▪
A "*" (asterisk) symbol indicates a channel that is connected to a detector that has
been zero calibrated successfully.
The number of detectors detected and calibrated is also displayed: