47
7.2.4
Jaw closing force
IT MAY BE POSSIBLE TO DAMAGE COLD PULL JAWS BY
INCORRECT SETTING OF THE CLOSING FORCE.
ENSURE THIS FORCE IS SET CORRECTLY AS BELOW.
Check the pressure regulators are set correctly (section 3.3.2).
Tweezer force is adjusted using the rotary control fitted on the cartridge.
The pressure will be stable once set and should not require adjustment.
If the jaw size is changed, or the pressure needs to be reset, proceed as follows: -
Turn the pressure control to zero (fully anticlockwise).
Operate the air valve to the jaws by pressing the or button.
Align the jaws over a ball of the size to be tested.
Slowly increase the air pressure (clockwise) until the jaws begin to close.
Increase the pressure until the jaws have closed around the ball, as shown below.
The jaws should slightly re-form the ball shape as shown, inspect by opening and closing the jaws
using the or button. These buttons will
R
R
Closing force
pressure gauge
Pressure adjuster