Load application parameter values related to leakage radiation:
According to standard NBR IEC 60601-1-3, the values for leakage radiation do not exceed the limit of 0.25 mGy
in one hour.
Conformity Statement
Dabi Atlante guarantees that the Spectro 70X x-ray machine, as concerns all models cited in this manual, conform
to standard NBR IEC 60601-1:1997
Radiation source set X SPECTRO 70X, NBR IEC 60601-2-28:2001
Troubleshooting
Overload Protectors
127V~ voltage
15 A~ circuit breaker
220V~ voltage
10 A~ circuit breaker
Quality equivalence filtration
Toshiba Tube
Kailong Tube
Tube glass
0.79 mm Al
0.85 mm Al
Isolating oil (oil column 10.5mm)
0.21 mm Al
0.21 mm Al
Acrylic
0.27 mm Al
0.27 mm Al
Additional filter (aluminum)
1.59 mm Al
1.59 mm Al
Total filtration
2.86 mm Al
2.92 mm Al
Filtration with quality equivalence in 70 kV and 2.5 mm of aluminum with a semi-reducing layer – according to
standard IEC 60522
Data for application of a 0.3-second load
Voltage
68 kVp ± 5kVp
Current
6.8 mA ± 1.2 mA
Power
0.34 kW
Current time product
2.1 mAs.
•
Maximum power applied intermittently for 1h = 288 KJ.
•
The x-ray tube voltage variation extension is of 1.4 kVp due to mains voltage fluctuation.
•
The field specified for radiation time conformity ranges from 0.3 to 2.5 seconds.
Faults
Causes
Solutions
Keys will not work
Equipment in unplugged
Plug it in
Fuse is burned
Replace fuse
Master switch is off
Turn switch on
Circuit breaker is off
Turn on circuit breaker
No power supply
Wait until supply is normalized
A semicircle appears on
Locating cylinder incorrectly
Apply the parallelism radiographic
radiographs
positioned
technique by using the auxiliary lines
of the collimating cylinder
Radiographs show a dark strip
Dark room
Light has penetrated the dark room.
Prevent sunlight or excessive illumination
from entering the dark room
Control shows a dark strip
Faulty electronic circuit
Unplug the equipment and call for
technical assistance
Completely dark radiographs
Excessive exposure time
Adjust exposure time (refer to “Radiographic
Techniques” in this manual)
Excessive developing time
In hot weather, the developer has quicker
action
Developer
Maybe the developer may contain
Note: Kodak developer does
insufficient mixture ( water/ developer
not use mixtures
ratio – 3:1)