Cleaning
Clean the equipment surface with a clean soft cloth
moistened in Aplic Odonto by Dabi Atlante or
another product with similar chemical properties.
The use of other chemical products is not
recommended as it may damage the
equipment.
To clean the fluid reservoir, remove it from the
product (fig. L). Clean it and put it back.
Remove the reservoir only when it is extremely
necessary.
Sterilization
Sterilize the sodium bicarbonate jet handpieces,
tips
,
tip
installation keys and the handpiece
protective cap by autoclave at up to 135 °C.
Remove the sodium bicarbonate jet
handpiecetip hose as it is not autoclavable.
Water drainage
After each day of use, place the equipment on a
dry piece of cloth, incline it sideways at an angle of
approximately 45°. Drain the air filter (fig. C, item
13) by pressing the depressurization pin until all the
water has been eliminated.
This procedure prevents clogging of the
sodium bicarbonate jet tip.
Maintenance
Preventive maintenance
In order to reduce equipment faults and
increasedurability, call Dabi Atlante Authorized
Technical Assistance and set a regular
preventivemaintenance plan.
Profi Class:
The system identifies pump wear
and, if necessary, suggests its replacement by
displaying a warning.
Corrective maintenance
In case the equipment presents any faults that are
not listed in “Troubleshooting”, contact Dabi Atlante
Authorized Technical Assistance.
Do not open the equipment and/or try to
repair it as such action may aggravate the
problem or even generate further
malfunctioning.
Circuit diagrams, list of parts, components,
others
In case they are necessary, ask Dabi Atlante for
electric diagrams and/or lists of parts, components
or other information, which will be supplied by
agreement.
Recommendations, care and
warnings
In case the sodium bicarbonate jet tip becomes
clogged, remove the hose and proceed with
unclogging. (fig. J).
Sodium bicarbonate dehydrates lips. Use some type
of moisturizer on the patient’s lips prior to
procedures.
Sodium bicarbonate use is contra-indicated for
patients with renal complications or who are being
submitted to hemodialysis.
The sodium bicarbonate powder spray may
interfere with respiratory disorders. In such cases,
medical advice is recommended.
The sodium bicarbonate jet handpiece must be
loosened at least once a week to prevent thread
locking.
The use of ultrasound is contraindicated for
patients and dentists wearing a pacemaker.
The scaler tips must be removed at least once aday
to prevent thread locking.
Profi Neo
: When the Endo Tips are used with
no cooling, the unit can be used for a
maximum of 4 minutes, followed by an 8-
minute waiting period, to prevent irreparable
damage to the ultrasound handpiece.
Procedures in case equipment
functioning is altered
In case the equipment presents any faults, firstly
check whether they are listed under
“Troubleshooting”. In case the problem cannot be
solved, turn off the equipment, unplug it from the
power source outlet and call Dabi Atlante
Authorized Technical Assistance.