Troubleshooting
Problem
What To Do
Deposits in the glass cylinder
of the lamp
It is quite normal for deposits to collect in the glass
cylinder of the lamp after the lamps have been
switched off. These deposits are the result of the
metal halide precipitating inside the lamp when it
cools down. They will disappear completely once
the lamp is switched on and warms up again.
The lamp does not ignite.
When troubleshooting the problem, please note in
certain, rare cases, it may take up to 2 minutes for a
lamp to ignite (such a lamp will ignite normally once
it has been ignited several times).
*Make sure that the main switch is on.
*Make sure the unit is properly plugged in. (Check
all plug connections.)
*Check all fuses in the unit and the building.
*If the lamp is defective, change the lamp.
*If the ignition unit is defective, contact the Daavlin
Service Department.
*If it is impossible to ignite the lamp or if the lamp
produces only a weak, bluish light, switch the device
off immediately to prevent damage to the ignition
units. Contact Daavlin Service Department.
Filters have slipped or are
defective
Never operate the unit if the filters have slipped or
are defective. Switch the device off. Contact
Daavlin Service Department.
16.8
Spare Parts and Accessories
Description
Daavlin Part No.
Metal Halide High Pressure Lamp, 1000 W
101MLLAMP
Colored Filter (260mm x 230mm)
100MLFLTRGLSRD
Clear Filter Filter (639mm x 292mm)
100MLFLTRGLSCLR
UVA-1 Irradiance Meter
925GOX97
Dose Patch (box of 100)
101MTDP100
Cleaning Wipes
908WPDISMNK
UV Safety Glasses
–
Fit Over Green
905GR
UV Safety Goggles with Elastic Band
–
Red
101RDG