This section describes the requirements for space and supplies in the building where
the Sterile Tube Fuser - Wet instrument will be installed.
Environmental and electrical power
requirements
The following general requirements must be fulfilled:
Parameter
Requirements
Allowed location
Indoor use only
Electrical power
100-120/220-240 V AC, 50/60 Hz
Transient overvoltages
Overvoltage category II
Operating temperature
2°C to 35°C
Humidity
20% to 80%, non-condensing
Altitude
Maximum 2000 m
Pollution degree
2
•
Make sure that the air flow to the instrument's ventilation fans is not blocked.
•
The instrument should not be exposed to sources of heat, such as direct sunlight.
•
Dust in the atmosphere should be kept to a minimum.
•
There should be at least 0.6 m of space around the instrument for easy access to the
whole equipment, both for normal operation and for maintenance and service oper-
ations.
•
Section 2.4 Danger zone, on page 25
) should be kept clear of
obstructions.
•
The instrument must be protected from liquid splashes.