A Blanking-plug is supplied for use: For safety we recommend
that the Blanking Plug is fitted at
all times
when there is no
cable attached to the Gauge body
Electrical connections to the Gauge
See pages 63, 64 & 65 for schematic connection details.
Power Supply
The Gauge requires a DC supply voltage between 7.5 and 30 volts.
There is an internal 1 amp fuse and crowbar circuit to protect the
instrument from over-voltage surges. The supply input is also
protected against accidental reverse polarity connection.
There is no ON/OFF switch. The Gauge will operate continuously
whilst power is connected. There will be an initial delay of about 2
seconds, and then the instrument will begin continuously
transmitting thickness-measurement data.
A “Test-Cable” is supplied in the main Gauge Kit, which can be
used to verify the Gauge operation. If you are using the Test-
Cable, connect Red to positive and Black to negative.
Data Output
The Gauge outputs its status and thickness measurements as RS-
422 serial data. The RS-422 output option can transmit the serial
data up to 1200 m (4,000 ft) along a twisted pair cable. The serial
Cygnus MINI ROV
Gauge Body
Blanking
Plug