104
To perform a basic inclination measurement:
1)
Make sure the computer running the MultiView application is within the range of the link module, about 30 ft
(10 m).
2)
In the MultiView window, for Right pane, choose Magnifier.
3)
Place the leveling wafer on the equipment you want to check. Align the leveling wafer so it matches the
MultiView display by doing the following.
•
Rotating the AMS until logo is right-side up.
•
Depending the model you are using, rotating the AMS so the notch is
pointing toward you
as you view the
AMS, or the flat side of the AMS is to the
left
as you view the AMS, as shown in the figure below.
Figure 5. Orienting the AMS
When you move the leveling wafer, the readings typically need some time to stabilize. The bubble displays and
numerical readouts are shaded during this stabilization period. For ALS1 systems, a message in the left bubble
display counts down the seconds (see the figure on the next page). Whenever the leveling wafer detects
movement, MultiView restarts the stabilization countdown. For ALS1 systems, you can set the length of the
stabilization period and the sensitivity that triggers the countdown (see “
Setting the Stabilization Criteria
” on
page
). For ALS2 and ALS2 Vertical wafers, MultiView determines the necessary stabilization time
automatically. You can’t set the stabilizati
on criteria for ALS2 or ALS2 Vertical wafers.
Notch