59
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Inspecting and Cleaning the External Surfaces of the Handpiece
Inspect and clean the handpiece housing, CoolView treatment guide, Dermastat standoff (if used), and
umbilical cable between patient use with a germicidal disposable wipe, such as Sani-Cloth
®
Plus or Sani-
Cloth
®
HB from Professional Disposables International, Inc.
1
Inspecting and Cleaning the Handpiece Lens(es)
The recessed handpiece lens(es) is (are) located at the distal end of the handpiece. Prior to treatment,
as well as periodically during treatment, inspect the lens(es) for cracks and debris. If the handpiece
lens(es) is (are) cracked or otherwise damaged, do not use the handpiece. If the handpiece lens(es) is
(are) dirty, clean as directed below.
WARNING
Before inspecting and cleaning the handpiece, ensure that the
system is either turned off or in STANDBY mode. Always wear
proper protective eyewear while the system is on. Severe eye
damage can occur in the event of unintended laser emission.
CAUTION
Do not allow liquid or cleaning solution to enter the handpiece
housing as damage may occur.
CAUTION
Do not attempt to sterilize the handpiece. Damage will occur if the
handpiece is autoclaved, immersed or otherwise handled
improperly.
1. Sani-Cloth Plus and Sani-Cloth HB are registered trademarks of Professional Disposables
International, Inc.
WARNING
The system must be turned OFF or in STANDBY mode prior to
inspecting the lens(es). Do not inspect the lens(es) while the system
is in READY mode. Always wear proper eyewear while the system
is ON. Severe eye damage can occur in the event of unintended
laser emission.
Summary of Contents for Excel V+
Page 1: ...I Operator Manual...
Page 8: ...viii EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 14: ...6 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 Remote Interlock Plug Footswitch...
Page 104: ...A 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 110: ...C 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...
Page 112: ...D 2 EXCEL V OPERATOR MANUAL D2325 REV A 2 19 THIS PAGE IS LEFT BLANK...