45
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
Physical Parameters
Size (W X D X H):
13 in X 18 in X 38 in (.33 m X .46 m X .96 m)
Weight:
140 lbs (63.5 kg)
Operating length of umbilical: 6.5 ft (2 m)
Minimum bend radius of
Storage
5 in (126 mm)
umbilical:
Momentary
2 in (50 mm)
Electrical Requirements
Voltage, Current
200 - 240 V~, 30 A
Frequency
50/60 Hz
Environmental Requirements
Temperature
Operating
5º C to 30º C
Non-operating -10º C to 50º C, coolant drained
Humidity
0 to 90%, non-condensing
Max. Heat Dissipation
4,000 Watts
Max. Altitude
Operating
15,000 ft
Non-operating 50,000 ft
Atmospheric Pressure
20 to 110 kPa
Maximum Allowable Temperature of Accessible Parts
Parts User/Patient Can Touch
Material
Duration
Limits
Enclosure
Plastic
> 1 min
48º C
Touchscreen
Plastic
> 1 min
48º C
Handpiece Window
Sapphire
> 1 min
20º C
Handpiece Treatment Guide
Metal
> 1 min
20º C
Classifications
FDA Classification
Class II Medical Device
CDRH Classification
Class IV Laser Product
European MDD 93/42/EEC
Laser Classification
Class 4 Laser to IEC 60825-1:2014
Type
BF
Applied Part
Handpiece