Electrical Utilities
The excel HR accepts mains power from 200-240 VAC. Power must be supplied from a dedicated
220±10% VAC, single-phase, 50/60 Hz source. The wiring should be rated for 30 amps (as per local
codes) “Hospital Grade” (NEMA L-20P or NEMA L-30P). Such a connection will ensure compliance with
allowable leakage current levels (<500 uA) per IEC 60601-1 for this device.
To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
System Specifications
Specifications subject to change without notice.
Treatment Beam
Type
Alexandrite
High Power, Long Pulse
Nd:YAG
Wavelength
755 nm
1064 nm
Fluence
4-100 J/cm
2
3-300 J/cm
2
Pulse Width
3 ms
0.1-300 ms
Repetition Rate
≤
2 Hz and single shot
≤
10 Hz and single shot
Delivery System
Permanently attached umbilical and handpiece
Treatment Spot
5–18 mm
3–18 mm
Epidermal Cooling
Full contact thermoelectric chiller
System Cooling
Self-contained, water to air heat exchanger
Aiming Beam
635 nm