Culligan® Premier Series Automatic Deionizer
44
Cat. No. 01024000
Test Communication Between Circuit Boards (TEST PROGFLOW)
The Test Progflow diagnostic verifies communication between the two systems in an alternating flow system.
SERVICE 1
STANDBY 2
FEB-01-13 12:11P
1. From the default home screen, press
to display the main menu.
3)SET DATE/TIME
4)ADV. SETUP
5)ACCESSORIES
>6)DIAGNOSTICS
2. Press
to select 6) DIAGNOSTICS, the diagnostics informa-
tion and settings.
1)MANUAL CONTROL
2)CHECK INPUTS
3)TEST PHONELIN
>4)TEST PROGFLOW
3. Press
to select 4)TEST PROGFLOW.
PROGRESSIVE TEST
Rx:0027
Tx:0027
4. The screen displays the communication status shown on the master system be-
tween the master and slave systems. Both numbers should be increasing in value
for the transmitter and receiver to verify that communication is working. Press any
key to view the next diagnostic screen. Refer to step 5 for the slave system screen
display.
PROGRESSIVE TEST
RCVD: 0000
5. The screen displays the communication status shown on the slave system be-
tween the master and slave system. This screen occurs at the same time as the
screen in step 4.
SERVICE 1
STANDBY 2
FEB-01-13 12:11P
6. The controller automatically returns to the home screen after a few moments.