System Setup
35
Cat. No. 01024000
SERVICE
FEB-01-13 12:11P
10.0GPM 161PPM
9. Press
until the deionizer controller displays the home screen.
SERVICE 1
STANDBY 2
FEB-01-13 12:11P
NOTE If the system type is ALTERNATING, the screen displays the status of
each deionizer. The tank that is not in service is either in STANDBY or
REGENERATING status.
Reset Date and Time
If, for some reason, the deionizer controller incorrectly displays the date and/or time, use the Set Date/Time setting to
change the setting.
NOTE If a modem is installed and connected to the phone line, the deionizer controller will check for the
correct time whenever the unit dials in.
1)INFORMATION
2)MANUAL REGEN
>3)SET DATE/TIME
4)ADV. SETUP
1. From the main menu screen, press
to 3) SET DATE/TIME.
2. Press
. The screen displays the Set Date/Time menu.
Use the procedure on page 30 to change the date and time settings. After changing the setting, press
to accept the
displayed screen value.
Date and Time Settings
Setting
Range
Unit of Measure
Default
SET MONTH
JAN–DEC
—
JAN
SET DAY
1–31
—
1
SET YEAR
2009–2040
—
2012
CLOCK TYPE
12 or 24
HR
12 HR
SET HOUR
12PM–12AM
or 00:00–23:59
AM/PM unless 24-hour
clock was selected
12PM
or 00:01
SET MINUTES
0–59
—
1
DAYLIGHT SAVING
YES or NO
—
YES
SERVICE 1
FEB-01-13 12:11P
10.0GPM 161PPM
3. Press
until the deionizer controller displays the home screen.
Menu Default (Re-Run First Time Setup)
This procedure will default the board to factory settings and initiate the first-time setup.
1. Power down the control.
2. While restoring power to the controller, press and hold down
and
for at least ten (10) seconds.
3. Release
and
.
The screen display will be blank—if the screen is not blank then repeat step 1.
4. Wait up to 30 seconds for the screen to light and then display FIRST TIME SETUP.
5. Follow the First Time Setup procedure.