5.2 Saving and Transferring Treatment Data
5.2.1 Device Usage
This device automatically saves the following treatment data:
•
ECG data
•
Usage information
The treatment data is automatically saved in the internal memory. This data can be transferred
to a personal computer (PC) and is not erased even if the device is turned off.
The i-PAD CU-SP2 saves the 3 most recent treatments and is able to save up to 17 hours for
each event. If more than 17 hours of ECG data are recorded for one event, any ECG data
over 17 hours will not be recorded.
When the device is used more than 3 times, it overwrites the oldest treatment data with the
newest data. Therefore, we recommend you to save the recorded treatment data by
transferring it to a PC after using the device.
If the battery pack is removed while the device is operating, the treatment data will not be
properly recorded. If you wish to remove the battery pack, turn the power off by pressing the
Power Button for more than 1 second, and then remove the battery pack.
5.2.2 Transferring Treatment Data
The treatment data may be transferred via an SD card or IrDA. The treatment data of all
patients recorded on the device is transferred using the SD card method, whereas the treatment
data of one patient is selectively transferred with the IrDA method.
1. Copying Treatment Data by Using an SD Card
①
Format the SD card on the PC to the FAT (FAT16) format.
②
Open the SD card cover on the device and insert an SD card into the port.