background image

Installation and Operation Manual

GG-H2S  

HYDROGEN SULFIDE GAS SENSOR

Summary of Contents for GG-H2S

Page 1: ...Installation and Operation Manual GG H2S HYDROGEN SULFIDE GAS SENSOR...

Page 2: ...ly in the manner described in this manual If the equipment is used in a manner not specified by Calibration Technologies the protection provided by the equipment may be impaired This equipment should...

Page 3: ...es 920 N Tradewinds Pkwy Columbia MO 65201 866 394 5861 sales ctiengineering com www ctiengineering com General description 4 Installation 4 Locating the sensor 4 Installation guidelines 5 Wiring 6 Op...

Page 4: ...oisture variations that occur during daily process cycles The transmitter circuit board is sealed in potting compound protecting sensitive electronic components and copper tracing from corrosion The s...

Page 5: ...om mercury vapor lights variable speed drives and radio repeaters Protect sensor from physical damage forklifts etc If mounting on a wall with studs the mounting screws should be screwed into the stud...

Page 6: ...scale range Wiring Guidelines Always use 3 conductor insulate stranded shielded copper cable Do not pull sensor wiring with AC power cables This can cause electrical interference If cable runs cannot...

Page 7: ...in mA Always use mVDC or VDC voltmeter settings Zero Calibration After the unit is installed and has been powered up for a minimum of 12 hours the unit can be zero calibrated by the following Be sure...

Page 8: ...ed Fault red LED On steady if supply voltage is less than 10VDC Status amber LED Blinks once per second if RFI is detected Power green LED On steady to Indicate Power Blinks once per second for 60 sec...

Page 9: ...installed in the same room All tests and calibrations must be logged Always disconnect power at the controller before performing any wiring at the sensor Sensor Life These electrochemical cells are ex...

Page 10: ...Linearity 0 5 of full scale Repeatability 1 of full scale Temperature Range 0 F to 120 F 18 C to 49 C Humidity Range 5 to 100 condensing Dimensions 7 7 high x 6 7 wide x 3 8 deep Weight 3 lbs Wiring C...

Page 11: ...f the product by and person other than an authorized dealer or contractor or the installation of unapproved parts on the product The obligations set forth in this warranty are conditional on a proper...

Page 12: ...ctiengineering com 866 394 5861 GG H2S DOC1 0 20190605...