CSI PRD-SC30-125 Manual Download Page 3

English

3

• Reference 

Viperslide 

®

 Lubricant product IFU for 

product handling. 

11. Equipment, Setup, and Test

11.1. Equipment

In addition to the OAS components, equip the 

operating room with the following:

• Introducer, guide sheath, or guide catheter - see 

Appendix B for sizing recommendations

• 

Standard IV pole with five wheels and a 20 inch 

diameter base

• 1000 mL bag of normal saline

• Fluoroscopic imaging equipment

• Standard hospital grade, electrical wall outlet

• Other equipment, as needed, for interventional 

procedures

11.2. OAS Pump Set Up

1.  Use the IV pole screw clamp to attach the 

OAS pump to a standard IV pole making sure 

to attach the OAS pump to the IV pole at a 

distance not greater than 60 in (153 cm) from 

the floor to the top edge of the OAS pump.

2.  Hang the low saline level sensor and cord, by 

the closed loop, from the horizontal arm of the 

standard IV pole.

3.  Plug the low saline level sensor connector into 

the back of the OAS pump (Figure 5).

Figure 5. 

Plug in the low saline level sensor

4.  Verify that the power cord is connected to the 

back of the OAS pump.

5.  Insert the other end of the power cord into the 

electrical wall outlet.

Warning:

 To avoid risk of electric shock, this 

equipment must only be connected to a supply 

mains with protective earth.

Warning:

 Ensure the power cord connection 

to the OAS pump and the on/off switch is 

accessible at all times.

Caution:

 Do 

not

 allow fluid to leak onto 

electrical connections of the OAS pump.

11.3. Preparing the Bag of Saline and 

Lubricant 

Ensure that the OAS pump is powered off by 

pressing the 

Master Power

 switch on the back 

of the OAS pump to off and ensure that no LEDs 

are illuminated on the OAS pump control panel 

(Figure 6). 

Figure 6. 

OAS Pump control panel

A.  Low saline red LED indicator

B.  Prime button

C.  Start button and green LED indicator

D.  Status yellow LED indicator
1.  Prepare a full 1000 mL bag of normal 

saline solution with lubricant. Refer to the 

ViperSlide 

®

 Lubricant Instructions for Use for 

lubricant preparation instructions.

2.  Hang the prepared saline bag with lubricant 

from the low saline level sensor on the 

standard IV pole.

Caution:

 Do 

not

 use glass bottles for the 

saline solution with lubricant or hang multiple 

saline bags from the low saline level sensor 

as this will disable the Low Saline Information 

signal.

11.4. Connecting the OAD to the OAS Pump

Remove the sterile saline tubing from the OAD 

package and pass the saline bag spike end of the 

saline tubing out of the sterile field. Connect the 

other end of the saline tubing luer to the device 

luer. Additionally, pass the OAD power cord out of 

the sterile field. 

Perform the following:
1.  Connect the saline tubing to the saline bag 

with lubricant using standard institution 

procedures.

2.  Open the door, located on the front of the OAS 

pump, by rotating the door in the direction of 

the arrow (Figure 7).

3.  Place the saline tubing over the pump rollers 

so that the tubing positioners align with the top 

and bottom V-guides on the pump (Figure 7).

Figure 7. 

Placing the saline tubing within the OAS 

Pump

A.  Saline tubing positioners

B.  Saline tubing

C.  V-guides

D.  OAS Pump door
4.  While closing the door, verify that there is no 

pinching of the saline tubing and ensure that 

there is slack in the saline tubing between the 

OAS pump and saline bag with lubricant.

5.  Verify that the saline tubing is properly 

inserted into the saline tubing V-guides and 

that there are no kinks or damage to the saline 

tubing.

6.  Press the 

Master Power 

switch, on the back 

of the OAS pump, and verify that the red or 

yellow LED is illuminated on the OAS pump 

control panel.

7.  Connect the OAD power cord to the OAS 

pump.

8.  Remove the driveshaft from the dispenser coil.

9.  Purge air from the OAD and the saline tubing 

as follows:

a.  Verify that the saline tubing is connected 

to the OAD.

b.  Press the green 

Start

 button on the OAS 

pump control panel to start saline flowing 

through the saline tubing. Verify that the 

green LED illuminates.

c.  Press and hold the 

Prime

 button on the 

OAS pump control panel to purge air from 

the saline tubing. Continually pressing the 

Prime

 button will pump saline through the 

tubing at an increasing flow rate. Releasing 

the 

Prime

 button will decrease the flow to 

the low flow rate after two seconds.

d.  Verify that saline is exiting from the OAD 

sheath near the crown.

e.  Continue priming to ensure there are no air 

bubbles within the saline tubing and use 

standard hospital procedures to aspirate or 

purge air from the lines.

f.  After verifying there are no air bubbles 

within the saline tubing, discontinue 

priming.

Warning:

 Do 

not

 attempt aspiration 

through the OAD or saline line while 

placed within the body. If saline is pulled 

out through the OAD or saline line, air may 

enter the system.

11.5. Testing the OAD

11.5.1 Testing OAD Crown Advancement

Before inserting any portion of the OAD into the 

body, ensure that axial movement of the OAD 

crown advancer knob will produce smooth travel 

of the crown.

Caution:

 Do 

not

 spin the crown during this test.

1.  Ensure that the crown advancer knob is in 

the unlocked position as this will allow free 

axial travel of the crown advancer knob.

2.  While visually monitoring the crown, slowly 

move the crown advancer knob in a back 

and forth motion. The maximum travel 

of the crown advancer knob, and the 

corresponding maximum travel of the shaft 

tip, is 5.9 inches (15 cm).

11.5.2 Optional: Testing OAD Crown Rotation

This test is optional, but is performed 

chronologically after testing crown 

advancement.

Note:

 Hold the guide wire firmly during the test. 

When the test is complete, the OAD is ready for 

use and the guide wire can be inserted through 

the introducer, guide sheath, or guide catheter.

1.  Push the crown advancer knob fully 

proximal, away from the nose of the handle, 

and release the guide wire brake before 

threading the guide wire through the OAD 

drive shaft.

2.  Grasp the proximal end of the guide wire 

and thread the guide wire through the 

opening in the OAD drive shaft distal tip.

Caution:

 Do 

not

 operate the OAD if there is 

a bend, kink, or tight loop in the guide wire. 

A bend, kink, or tight loop in the guide wire 

may cause damage to and malfunctioning of 

the OAD during use.

3.  Continue feeding the guide wire into the 

OAD drive shaft until the guide wire appears 

at the rear of the OAD.

4.  Lock the guide wire in place by pressing 

down on the guide wire brake lever as the 

crown will not spin if the guide wire brake is 

unlocked.

5. 

Verify that saline is still flowing freely out of 

the saline sheath tip. Verify that the saline 

tubing is properly connected to the saline 

bag, that the saline tubing routes correctly 

through the saline tubing guides, and that 

the saline tubing is properly connected to 

the OAD.

6.  Hold the OAD sheath a few centimeters from 

the crown while making sure that the crown 

is not in contact with any objects. Verify 

there is no pinching of the OAD sheath at 

any time during OAD operation. 

7.  Press and release the 

On/Off

 button located 

on top of the crown advancer knob to 

activate crown rotation. The OAD is preset 

to low speed, and the illuminated LED on the 

OAD will indicate that the OAD is operating 

at low speed.

8. 

Check that the flow of saline is increasing 

and that the shaft and crown are beginning 

to spin.

9.  Immediately press and release the 

On/Off

 

button to stop the shaft and crown from 

spinning and to complete the test.

11.6. Initiating the Atherectomy Procedure

1. 

Gain vessel access using the physician’s 

preferred methodology.

2.  Access the treatment site with an 

appropriately sized introducer, guide sheath, 

or guide catheter.

Note:

 For radial access, use a preferred guide 

catheter or guide sheath of an appropriate 

length.

3.  Use angiography to locate, visualize, and 

evaluate the lesion.

4. 

If desired, use the thumb and index finger to 

gently impart a slight curve or J-shape to the 

distal spring tip of the 

ViperWire 

®

 Peripheral 

Guide Wire.

5.  If use of the guide wire torquer is desired, 

attach the torquer to the guide wire by holding 

the distal end of the torquer and rotating the 

proximal end counterclockwise to tighten.

6.  Approach and cross the lesion, with the 

ViperWire 

®

 Peripheral Guide Wire, using the 

physician’s preferred methodology.

12. OAS Directions for Use

12.1. Performing the Atherectomy Procedure

1.  Ensure that the OAD guide wire brake lever is 

open (in the up position).

2.  Lock the crown advancer knob at 1 cm from 

the fully proximal position.

3.  While keeping guide wire placement 

stationary, advance the OAD drive shaft over 

the 

ViperWire 

®

 Peripheral Guide Wire and 

through the hemostasis valve.

4.  Under direct visualization, gently advance the 

OAD crown over the 

ViperWire 

®

 Guide Wire 

to a position approximately 1 cm proximal to 

the lesion. Verify that the OAD distal tip is not 

within the lesion when the crown and drive 

shaft begin to spin.

(Optional)

 Use the 

GlideAssist 

®

 mode feature 

to facilitate advancing the OAD crown over the 

ViperWire 

®

 peripheral guide wire. This feature 

is only available on 1.25mm Micro OAD.

a.  Enable 

GlideAssist 

®

 mode by pressing 

and holding the low speed button. Release 

the button once the low speed light begins 

to slowly blink. The slowly blinking light 

indicates 

GlideAssist 

®

 mode is enabled.

b.  Ensure the 

guide wire is secure

 by 

locking the guide wire brake or by holding 

the guide wire with either fingers or the 

guide wire torquer.

c. 

Press and release the On/Off button on 

top of the crown advancer knob to activate 

crown rotation. The low speed light 

will rapidly blink indicating the crown is 

spinning in 

GlideAssist 

®

 mode.

d.  Stop the OAD crown rotation by pressing 

and releasing the On/Off button on top of 

the crown advancer knob. The low speed 

light will slowly blink indicating the OAD 

is no longer spinning but continues to be 

in i mode.

e.  Disable 

GlideAssist 

®

 mode by pressing 

and immediately releasing any speed 

button while the crown is not spinning. 

The low speed light will stop blinking, yet 

remains illuminated indicating the OAD is 

now in treatment mode.

Warning:

 Spinning the crown using 

GlideAssist 

®

 mode can be done with 

the OAD guide wire brake lever in either 

the locked or unlocked position. If using 

GlideAssist 

®

 mode with the guide wire 

brake in the unlocked position, the 

guide 

wire must be held

 using either fingers or 

the guide wire torquer. If using the guide 

wire torquer, ensure that it is securely 

fastened to the guide wire before starting 

to spin using GlideAssist 

®

 mode.

Note:

 If the brake configuration is changed 

from either the locked or unlocked position 

while spinning in GlideAssist mode, the 

crown will automatically stop spinning yet 

the OAD will remain in GlideAssist mode.

5.  Inject contrast medium through a port in the 

hemostasis valve to verify that the size of the 

crown is compatible with the treatment area 

diameter (see Appendix C).

6.  Verify that the guide wire spring tip is distal to 

the lesion and is not in danger of coming in 

contact with the advancement of the spinning 

crown and drive shaft tip.

7.  Push down on the guide wire brake lever to 

engage the guide wire brake. The crown will 

not spin if the guide wire brake is not locked.

8.  Unlock and move the crown advancer knob 

to the fully proximal position to relieve any 

compression in the driveshaft.

9.  Press and release the 

On/Off

 button on top 

of the crown advancer knob to activate crown 

rotation. The OAD is preset to low speed, and 

the illuminated LED on the OAD will indicate 

that the device is operating at low speed.

10. Audibly verify that the OAD drive shaft and 

crown are orbiting at a stable speed.

11. Slowly advance the crown advancer knob to 

begin atherectomy of the lesion at a travel 

rate between 1 mm per second and 10 mm 

per second. Using imaging, continually verify 

that the crown and the crown advancer knob 

are moving 1:1 with one another. Ensure that 

the OAD handle remains horizontal during the 

procedure to minimize saline leakage from 

the OAD.

12. Using a series of intermittent treatment 

intervals and rest periods, slide the crown 

advancer knob to move the crown back and 

forth across the lesion always returning to the 

proximal side of the lesion when the interval 

set is complete.

Warning:

 Once the OAD has reached full 

speed (as indicated by a stable pitch), do 

not

 allow the orbiting crown to remain in one 

location as it may lead to vessel damage. 

Continue to maintain a travel rate between 

1 mm per second and 10 mm per second.

A rest period of 30 seconds, is recommended 

for every 30 seconds of treatment, with a 

maximum treatment time of 8 minutes per 

OAD. The OAS pump will emit a beep after 

every 25 second interval of treatment time. 

Use contrast injections through the introducer, 

guide sheath, or guide catheter only during rest 

periods to fluoroscopically evaluate results.

Warning:

 

Maximum total treatment time 

should not exceed 8 minutes per OAD.

 If 

maximum total treatment time is exceeded, 

the OAD shaft, crown, and ViperWire 

®

 

Peripheral Guide Wire may begin to 

exhibit signs of wear and result in a device 

malfunction and possible injury to the patient. 

13. Evaluate the reduction of the stenosis. 

14. If reduction of the stenosis is not adequate, 

perform one of the following:

• Continue to treat the lesion by moving the 

crown back and forth across the lesion per 

the instructions above. 

• Increase the rotational speed of the crown 

by using the crown rotation speed buttons 

on the handle of the device.

15. 

Perform a final angiogram. 

12.1.1 Replacing the Bag of Saline and 

Lubricant

The low saline level sensor triggers an audible 

information signal every 5 seconds, for a total 

of 30 seconds, if there is less than 200 mL 

(± 100 mL) remaining in the bag of saline and 

lubricant during a treatment period. If the low 

saline level sensor triggers during a rest period, 

only the red low saline LED is illuminated. 

Perform the following to replace the bag of 

saline and lubricant:

1.  Ensure that the OAS pump is stopped by 

pressing the green 

Start

 button on the OAS 

pump control panel and verify that the green 

LED, on the OAS pump control panel, is not 

illuminated.

2.  Prepare a new 1000 mL bag of normal 

saline solution with lubricant. Refer to the 

ViperSlide 

®

 Lubricant Instructions for Use for 

lubricant preparation instructions.

3.  Remove the low bag of saline and lubricant 

from the low saline level sensor on the IV 

pole.

Summary of Contents for PRD-SC30-125

Page 1: ... D and E for orbit performance for the different crowns Table 1 Micro Crowns Model Number Crown Size mm Nose Length mm Shaft Length mm PRD SC30 MICRO 1 25 7 145 Nose length is the length of the drive shaft from the crown to the distal tip of the shaft Table 2 Solid Crowns Model Number Crown Size mm Nose Length mm Shaft Length mm PRD SC30 125 1 25 7 145 PRD SC30 150 1 50 10 145 PRD SC30 200 2 00 30...

Page 2: ...own to the point of contact with the guide wire spring tip Distal spring tip detachment and embolization may result Make sure there is a minimum of 10 cm between the guide wire spring tip and the distal end of the shaft Always advance the orbiting abrasive crown by using the crown advancer knob Never advance the crown by advancing the drive shaft or OAD handle Guide wire buckling may occur and per...

Page 3: ...he OAD is preset to low speed and the illuminated LED on the OAD will indicate that the OAD is operating at low speed 8 Check that the flow of saline is increasing and that the shaft and crown are beginning to spin 9 Immediately press and release the On Off button to stop the shaft and crown from spinning and to complete the test 11 6 Initiating the Atherectomy Procedure 1 Gain vessel access using...

Page 4: ... immerse the OAS pump into fluids Do not use solvents or abrasive cleaners to clean the OAS pump as these may damage the OAS pump and OAS pump components Caution Completely dry the OAS pump before reconnecting the OAS pump to wall power and powering on the OAS pump Clean the OAS pump immediately after each use by following the steps below 1 Prepare an enzymatic detergent such as Enzol per manufact...

Page 5: ...on has authority to bind CSI to any representation warranty or liability except as set forth in this Disclaimer of Warranty CSI may at its sole discretion replace any device that is determined to have been out of specification at the time of shipment The exclusions disclaimers and limitations set forth in this Disclaimer of Warranty are not intended to and shall not be construed as to contravene m...

Page 6: ... Table B2 Solid Crown Crown Diameter mm Model Number Orbital Atherectomy Device Maximum Outer Diameter mm inches Minimum Introducer or Guide Sheath Internal Diameter French inches Guide Catheter Sizing 1 25 PRD SC30 125 1 80 0 070 6 0 079 See guide catheter manufacturer specifications for lumen diameter 1 50 PRD SC30 150 1 80 0 070 6 0 079 2 00 PRD SC30 200 2 00 0 079 6 0 079 1 25 RAD SC30 125 1 6...

Page 7: ... High Speed 120 krpm 1 3 1 5 1 7 1 9 2 1 2 3 2 5 2 7 2 9 0 2 4 6 8 10 12 14 16 18 20 ORBIT DIAMETER MM NUMBER OF PASSES 1 25 MM SOLID CROWN 200 CM LENGTH ORBIT RESULTS 60k RPM Low Speed 90k RPM Medium Speed 120k RPM High Speed 1 50 1 60 1 70 1 80 1 90 2 00 2 10 2 20 0 2 4 6 8 10 12 14 16 18 20 ORBIT DIAMETER MM NUMBER OF PASSES 1 50 MM CLASSIC CROWN 145 CM LENGTH ORBIT RESULTS Low Speed 60 krpm Me...

Page 8: ...ed 2 00 2 10 2 20 2 30 2 40 2 50 2 60 2 70 2 80 2 90 0 2 4 6 8 10 12 14 16 18 20 ORBIT DIAMETER MM NUMBER OF PASSES 2 00 MM CLASSIC CROWN 145 CM LENGTH ORBIT RESULTS Low Speed 60 krpm Medium Speed 90 krpm High Speed 140 krpm 2 2 5 3 3 5 4 4 5 5 0 2 4 6 8 10 12 14 16 18 20 ORBIT DIAMETER MM NUMBER OF PASSES 2 00 MM SOLID CROWN 145 CM LENGTH ORBIT RESULTS Low Speed 60 krpm Medium Speed 90 krpm High ...

Page 9: ...s tailles de couronne Voir les annexes D et E pour la performance rotationnelle des différentes couronnes Tableau 1 Micro couronnes Numéro de référence Taille de la couronne mm Longueur de l embout mm Longueur de l arbre mm PRD SC30 MICRO 1 25 7 145 La longueur de l embout correspond à la longueur de l arbre d entraînement de la couronne jusqu à la pointe distale de l arbre Tableau 2 Couronnes ple...

Page 10: ...te pendant que la couronne de l OAD tourne Cela pourrait entraîner la défaillance de l OAD ou des lésions chez le patient Ne pas tenter d aspirer par l OAD ou la tubulure de sérum physiologique quand il est placé dans le corps Si du sérum physiologique est retiré de l OAD ou de la tubulure de sérum physiologique de l air peut entrer dans le système Si de l air est détecté dans le système pendant q...

Page 11: ...que et s assurer que la tubulure n est pas tendue entre la pompe de l OAS et la poche de sérum physiologique contenant du lubrifiant 5 Vérifier que la tubulure à sérum physiologique est correctement insérée dans les guides en V destinés à recevoir la tubulure et que celle ci n est pas pliée ou endommagée 6 Appuyer sur l interrupteur d alimentation principal au dos de la pompe de l OAS et vérifier ...

Page 12: ...ivantes 1 Arrêter la rotation de la couronne et de l arbre d entraînement en appuyant sur le bouton marche arrêt situé sur le dessus du bouton d avancement de la couronne et en le relâchant 2 Débrancher le cordon d alimentation de l OAD de la pompe de l OAS 3 Laisser l introducteur la gaine guide le cathéter guide et le guide en place relâcher le frein du guide sur l OAD et reculer la gaine et l a...

Page 13: ...ssitent des précautions spéciales concernant la compatibilité électromagnétique CEM Installer et utiliser les appareils électromédicaux conformément aux informations de CEM suivantes Ne pas placer d équipements de communication radiofréquence RF portables ou mobiles à proximité immédiate des appareils électromédicaux car ils peuvent en perturber le fonctionnement S assurer que les champs magnétiqu...

Page 14: ...ière à être exécutoire dans toute la mesure du possible Si la partie concernée ne peut pas être modifiée cette partie peut être coupée et les autres parties de la présente exonération de garantie resteront pleinement en vigueur Annexe A Dépannage du système Si dans chacune des situations ci dessous un problème de la pompe de l OAS ou de l OAD ne peut pas être résolu remplacer le composant recomman...

Page 15: ...er l OAD dans un vaisseau dont la taille est trop petite pour la couronne Le diamètre du vaisseau de référence au niveau du site à traiter doit être d au moins 2 00 mm pour la micro couronne de 1 25 mm Remarque Un passage est défini comme un va et vient à travers la lésion Les données sur l orbite qui sont présentées sont fondées sur une distance de passage de 6 cm à une vitesse de déplacement de ...

Page 16: ...RPM Haute vitesse 1 75 2 25 2 75 3 25 3 75 4 25 4 75 0 2 4 6 8 10 12 14 16 18 20 DIAMÈTRE DE L ORBITE MM NOMBRE DE PASSAGES COURONNE PLEINE 1 75 MM LONGUEUR DE L ORBITE 180 CM RÉSULTATS 60 k RPM Basse vitesse 90 k RPM Moyenne vitesse 120 k RPM Haute vitesse 2 00 2 10 2 20 2 30 2 40 2 50 2 60 2 70 2 80 2 90 0 2 4 6 8 10 12 14 16 18 20 DIAMÈTRE DE L ORBITE MM NOMBRE DE PASSAGES COURONNE CLASSIQUE 2 ...

Reviews: