CSA Medical truFreeze Active Venting Setup Instructions Download Page 1

truF
reez
e

® 

A

ctiv
e V
enting Set

-up Instructions

1.  Position the console as 

desired for procedure and 
lock the wheels by pushing 
the red lever to either side.

2.  Remove the power cord from 

the holder and plug into 
standard wall power. Ensure 
the power cord is connected 
to the console.

 

Make sure the power cord 
is dry

.

3.  Power on the console using 

the green switch on the back 
of the device.

4.  Ensure that the red 

emergency stop button is off 
(NOT lit).

3

 

1

 

6

 

2

 

4

 

truF
reez
e

® 

A

ctiv
e V
enting Set

-up Instructions

16. 

If using the remote, ensure the remote control 
has two (2) working AA batteries installed by 
selecting the 

run 

tab and attempting to adjust 

the site counter. Reset site to “1”.

5.  The computer will run a series of built-in tests 

(BIT). Note: the device must have adequate 
liquid nitrogen to pass the BIT.

P

  See separate tank filling instructions.

 

6.  While the BIT is running, position the foot pedals 

on the floor as desired, ensuring the foot pedal 
cord is plugged into the console.

  Avoid touching the pins of the foot pedal 

connector.

5

 

CSA Medical, Inc.

For service, call: 866.481.7786

www.csamedical.com

© 2014 CSA Medical, Inc. All rights reserved.

truFreeze is a registered trademark of CSA Medical, Inc.

15-00292-00

Active Venting Set-up

 

Liquid nitrogen expands more than 700 times when changing to a gaseous state.  It is 
imperative for patient safety that adequate ventilation is ensured.

note:

 

some consoles also 

display a

 

shutdown 

button.

15.  To ensure the suction system has been set up 

correctly, depress the gray suction pedal and 
verify that the tubing draws suction. Then 
press the suction pedal again to turn off 
suction prior to the start of the procedure.

 

Please read the Instructions For Use and Operator’s Manual for complete 
information about the truFreeze System.