IRGASON
®
Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
25
TABLE 8-2. Multipliers and Offsets for Analog Outputs
Density
(mg m
–3
)
Voltage Output Multiplier
(mg m
–3
V
–1
)
Offset
(mg m
–3
)
CO
2
386.32
–102.59
H
2
O
8.65
–2.26
9. Maintenance
IRGASON operation requires six maintenance tasks:
•
routine maintenance
•
wick maintenance
•
analyzer window cleaning
•
zero and span
•
replacing the analyzer desiccant / scrubber bottles
•
factory recalibration
9.1 Routine Maintenance
Periodically do the following:
•
Check the humidity indicator card in the EC100 enclosure. If the highest
dot has turned pink, replace or recharge the desiccant bags. Replacement
desiccant may be purchased (pn 6714) or old desiccant packs may be
recharged by heating in an oven. See the manual
ENC10/12, ENC12/14,
ENC14/16, ENC16/18
, for more details
on recharging desiccant bags.
•
Make sure the LED Status Lights on the EC100 panel are green. If not,
verify that all connections are secure and that the instruments are powered.
Also check the individual diagnostic bits for the specific fault. See TABLE
Bits in the Sonic Diagnostic Flag
and TABLE
Bits in the Gas
Diagnostic Flag
.
9.2 Wick Maintenance
9.2.1 Sonic Wicks
The IRGASON cannot measure wind when water droplets completely obscure
the face of the sonic transducers. If the site experiences frequent rainfall, using
wicks on the sonic transducers may be desirable. The wicks help remove water.
Under some conditions, however, droplets may build up and remain on the
transducer faces. As soon as the water droplets evaporate or are removed, the
IRGASON will again measure wind. Remove the water droplets by dabbing a
cotton swab or tissue on the faces of the transducers. The user must use
discretion to determine if wicks are necessary. If wicks are not used, gently
remove the wicks from the sonic transducers, taking care not to damage or peal
the matching layer (rubber tips) from the brass housing of the transducers.
Remove the wicks during the winter. The wicks will accumulate snow or
freezing rain, resulting in blockage of the sonic signals. The CSAT3 Sonic
Wick Spares Kit (pn 28902) consists of three top wicks, three bottom wicks,
Summary of Contents for IRGASON
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 12: ...Table of Contents iv ...
Page 60: ...Appendix A Filter Bandwidth and Time Delay A 4 ...
Page 64: ...Appendix C Material Safety Data Sheets MSDS C 2 ...
Page 65: ...Appendix C Material Safety Data Sheets MSDS C 3 ...
Page 66: ...Appendix C Material Safety Data Sheets MSDS C 4 ...
Page 67: ...Appendix C Material Safety Data Sheets MSDS C 5 ...
Page 68: ...Appendix C Material Safety Data Sheets MSDS C 6 ...
Page 69: ...Appendix C Material Safety Data Sheets MSDS C 7 ...
Page 70: ...Appendix C Material Safety Data Sheets MSDS C 8 C 2 Decarbite MSDS ...
Page 71: ...Appendix C Material Safety Data Sheets MSDS C 9 ...
Page 72: ...Appendix C Material Safety Data Sheets MSDS C 10 ...
Page 73: ......