2.4 Test Result Status Indication
During the process of the test the LED will flash amber. If the test is passed I-Test will display the
following icon
and the LED
À
on the I-Test gas monitor pod will be illuminated green. If the test
is failed I-Test will display the following icon
and the LED on the I-Test gas monitor pod will be
illuminated red (see below).
Figure 8: Result indications
I-Test will also display the
icon if I-Test was unable to perform the configured test.
This may occur for example if a monitor is inserted with an unsupported sensor/gas fitment or if the
attached cylinder does not contain a suitable test gas for a specific sensor type in the monitor.
If the monitor is not correctly inserted in the gas monitor pod it may not be recognised by I-Test.
This will be indicated by the I-Test gas monitor pod LED red and the icon
flashing for a period of
approximately 30 seconds. The monitor must be removed from the I-Test and reinserted.
Note:
If a monitor is inserted into I-Test with a specific sensor due bump test (not all sensors may be due a
bump test on the same date) and the attached cylinder does not contain the required gas, I-Test will
display the
icon and not perform a bump test on any sensors.
If a monitor is inserted into I-Test with a specific sensor(s)
not
due a bump test and the attached
cylinder does not contain that required gas(s) but the cylinder does contain the gas for remaining
sensors due a bump test. I-Test will continue to bump test the sensors due and will not bump test the
sensor(s) not due.