115
10. Troubleshooting
Symptom
Cause
Action
No action is undertaken
when the monitor
is inserted and no
progress or result icons
appear on the screen.
The I-Test display is not showing
the home screen and is in a
functional menu.
USB to PC cable is still attached
to I-Test following configuration.
Navigate to the home screen,
remove the monitor, re-insert to
repeat test.
Remove USB cable from I-Test,
remove the monitor, re-insert to
repeat test.
I-Test Performs no test
when monitor is inserted
and display then shows
complete success icon
No test action has been selected.
Select a test, e.g. speedy bump,
via either I-Test Manager (see
) or via I-Test Front Panel
(see
), and then repeat
process.
I-Test Performs no test
when monitor is inserted
and display then shows
the
icon
The attached cylinder does not
contain the required gas for a
sensor within Gas-Pro that is due
a bump test.
The attached cylinder does not
contain the required gas for a
sensor within Gas-Pro that to be
calibrated.
Replaced the attached cylinder with
a cylinder containing the appropriate
test gases and reconfigure I-Test.
USB Flash Stick icon
is not available when
the USB Flash Stick is
inserted.
USB functionality is not available
when I-Test is operating on
backup battery.
Connect the power adapter to a
suitable mains outlet and connect
its DC connector to the power input
socket on the I-Test.
I-Test does not respond
when power switch is
operated to turn on the
unit.
DC power supply is not
connected and backup battery is
exhausted.
USB to PC cable is still attached
to I-Test following configuration.
Connect DC power supply and
power cycle I-Test
Remove USB cable from I-Test and
power cycle I-Test.
Contents
Exit
Unpacking
Set-up
Operating I-Test
I-Test
Manager
Configuring
I-Test
Warranty
Overview
Contacts
Appendices
Troubleshooting
Specification