4.4.3 Calibration
I-Test offers three calibration options dependent upon user requirements, Standalone Calibration,
Calibration after Bump Test Fail and Calibrate if expired (refer to
Standalone calibration and calibration after bump test fail are available via the front panel, calibrate if
expired is available via I-Test Manager.
4.4.3.1 Standalone Calibration
This option should be selected if I-Test is to be configured to undertake only a calibration function,
when a gas monitor inserted into I-Test.
Standalone calibration will calibrate all sensors in the gas monitor; this will occur even if the sensor
was not due a calibration.
Standalone calibration can
not
be selected when speedy bump is selected.
If standalone calibration is selected with smart bump this will change the functionality of the
calibration process to that as described below in Bump Test Fail Calibration.
Standalone calibration can be selected with download gas monitor log files (see
X
To select a Standalone Calibration from the Test Sequence screen use the arrow keys to navigate
to the Calibration icon
then press
OK
to highlight the Calibration icon
À
.
X
A box will now appear around the icon indicating that Standalone Calibration function has been
selected and this test will be performed if a gas monitor is inserted into the I-Test.
i
To ensure a standalone calibration is undertaken, speedy bump and smart bump must not
be selected
X
Press
to return to Test Settings screen and then press
again to return to main menu screen.
Note: Smart bump functionality is only applicable to Gas-Pro, a T4 I-Test test screen will not display
the smart bump icon
.
Figure 83: Standalone Calibration screen