DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-25
Step
Description
Images illustration
6
Press on [Auto Options] button on
Main Menu screen.
Auto Options screen as shown will
prompt.
User can disable some of the
components such as heater or pump
if it is not in use on the following
process.
Please refer to section
5.8 Auto
Options Screen
for detail
description
.
Press on [Main Menu] button to go
back to Main Menu screen.
7
Logout form Engineer level when all
the settings are complete.
N/A
4.8.2
Rinse Station Auto Cycle Cleaning Process
NOTE
There is no interlock between tank temperature set point and
[CYCLE START]
button. Hence, always heat up tank to
process temperature set point prior to triggering
[CYCLE
START]
button to begin cleaning process.
NOTE
Liquid level must be above low liquid level sensor to trigger
heater, recirculation pump and ultrasonic to turn ON.
Operation stops if low liquid level is detected.
NOTE
Press
[EMO]
button at control panel to turn OFF the system. If
user wants to turn ON the system, reset the
[EMO]
button
and press [SYSTEM ON] button in touch screen Console
Power screen.