DMS-R-IM
Interpreting HMI
April 7, 2021
Digital Modular System - Rinse
5-9
Parameter
Description
Dwell Time
The length of time a basket remains in a station
before it proceeds to the next process or station
Note
: Allowable setting range:
Hour: 0 ~ 8 hour
Minute: 0 ~ 59 minute
Second: 0 ~ 59 second
Sonic Time
The length of time for ultrasonic to turn ON during
cleaning process. The ultrasonic ON time and process
dwell time begin simultaneously.
Ultrasonic time must be shorter than dwell time.
Note
: Allowable setting range:
Hour: 0 ~ 8 hour
Minute: 0 ~ 59 minute
Second: 0 ~ 59 second
Sonic Power
The desire ultrasonic generator power (%) during
Auto
cleaning process (when
[CYCLE START]
button is pressed.