background image

User Manual for Handheld Pulse Oximeter

I

Instructions to User

Dear Customer,
Thank you for purchasing this quality product. Please read the
manual very carefully before using this device. Failure to follow
these instructions can cause measuring abnormality or damage to
the Oximeter.
No part of this manual may be photocopied, reproduced or
translated into another language without the prior written consent.
We reserve the right to revise and amend it at any time without
prior notice.
Version of the Manual: Ver 1.0
Issued Date: October 28

th

, 2016

All rights reserved.

Notes:

The contents contained in this manual are subject to change
without prior notice.

Information furnished by

Creative

is believed to be accurate

and reliable. However, no responsibility is assumed by

Creative

for its use, or any infringements of patents or other rights of
third parties that may result from its use.

3502-2290079

Summary of Contents for OXY-110

Page 1: ...ed into another language without the prior written consent We reserve the right to revise and amend it at any time without prior notice Version of the Manual Ver 1 0 Issued Date October 28th 2016 All...

Page 2: ...NOT use the oximeter while the Patient is under MRI or CT scanning This device is NOT MRI Compatible Warnings Discomfort or pain may occur if using the sensor of this device continuously on the same...

Page 3: ...ation corrosive substances explosive materials high temperature and moisture If the Oximeter gets wet please stop operating it and do not resume operation until it is dry and checked for correct opera...

Page 4: ...when enclosure tilted up to 15 Please pay attention to the effects of lint dust light including sunlight etc Declaration of Conformity The manufacturer hereby declares that this device complies with...

Page 5: ...ent 7 3 1 SpO2 Measurement 7 3 2 Temperature Measurement optional 9 4 Operation 11 4 1 Power on off the Oximeter 11 4 2 Default Display Screen 11 4 3 Menu 14 4 4 Record 27 5 Technical Specifications 3...

Page 6: ...anual for Handheld Pulse Oximeter 1 1 Overview 1 1 Appearance Power Left Mute Right Display screen Figure 1 1 Front View Rotate lock Up Setting Down Alert indicator Menu Confirm Record Back Power indi...

Page 7: ...sound On menu or sub menu screen short time press it to move the cursor right or adjust the parameter values Note when the device alert indication is on short time pressing will mute the alert sound...

Page 8: ...r disconnected or the measured value exceeds the preset alert limit value then the alert indicator will keep orange flashing 9 Power saving mode indicator If the device is set as power saving mode the...

Page 9: ...Real time battery status display and low battery voltage indication Auto power off is available Audible and visual alert function is available Data uploading to PC for management Optional Power saving...

Page 10: ...held Pulse Oximeter 5 2 Preparation 2 1 Charger Base Connection Put the device into the charger base properly and note the polarity markings as shown in figure 2 1 Figure 2 1 Charger base connection O...

Page 11: ...ly is 100 240VAC with 50 60Hz 3 External power supply from the USB cable Use the USB cable with micro USB connector connect one end of the USB cable to the connector on the device marked with or to th...

Page 12: ...e SpO2 probe indicates a successful connection 3 Insert one finger index finger is preferred the nail should be not too long into the clip of the probe according to the finger mark as shown in figure...

Page 13: ...e tissue Do NOT put the SpO2 probe and pressure cuff on the same limb otherwise the blood pressure measurement may affect the SpO2 measurement The device is calibrated to display functional oxygen sat...

Page 14: ...ay cause damage to the probe The infrared temperature probe is as shown in figure 3 2 Please place the infrared temperature probe in a stable ambient temperature for 30 minutes before taking a measure...

Page 15: ...frared temperature probe will switch to standby state automatically if there is no operation for 1 minute If a further measurement is needed press the measuring key and repeat step 2 and step 3 Normal...

Page 16: ...Long pressing Power Left key for 1 2 seconds then the oximeter will be powered on The oximeter will do self test and then show the software version as shown in figure 4 1 refer to your oximeter for ac...

Page 17: ...nd appears simultaneously Alert sound is SpO2 mark SpO2 value Pulse Plethysmogram Current time SpO2 high low limit setting range Patient type Sound indicator Battery indicator Pulse rate mark and unit...

Page 18: ...e lock icon turns to blue it means the auto rotation function is turned on if you place this oximeter horizontally then the display shows in horizontal as shown in figure 4 2B Sound indicator means th...

Page 19: ...rent corresponding storing space is not full If the memory is full the data storing will continue in such way the new record will overwrite the oldest record so that it s recommended to upload the sto...

Page 20: ...grey and the patient type on upper left corner turns to pink Measurement mode Select the measurement mode Spot check mode and Continuous mode for option SpO2 record Recall and review the records store...

Page 21: ...ode etc see Section 4 3 7 for details Alert Set the low alert limit for SpO2 and the high low alert limit for HR see Section 4 3 8 for details Help To view the tips information of SpO2 measurement and...

Page 22: ...an edit the User ID and move the cursor on OK to confirm the edit the edit screen is as shown in figure 4 4B Figure 4 4B User ID edit screen 4 3 2 User On main menu screen move the cursor on User and...

Page 23: ...ement Mode Setup screen as shown in figure 4 6 Figure 4 6 Measurement mode setup screen 4 3 4 SpO2 Record On main menu screen move the cursor on SpO2 Record and press Confirm key then the oximeter ent...

Page 24: ...rd and press Confirm key then the oximeter enters into temperature record list screen as shown in figure 4 8 Figure 4 8 TEMP record list screen 4 3 6 Date On main menu screen move the cursor on Date a...

Page 25: ...key to confirm and exit from date setting 4 The procedures of adjusting Month Day Hour Minute and Second value are the same with Year adjustment Date Format DD YY MM Time Format HH MM SS Note The sett...

Page 26: ...figure 4 10B Language This oximeter provides the display with two languages English and Simplified Chinese the factory default is English as shown in figure 4 10C Auto power off To turn on off the Au...

Page 27: ...figure 4 10H Version For viewing version number of the software as shown in figure 4 10I Demo Enter into the Demonstration mode as shown in figure 4 10J Figure 4 10A Brightness setup Figure 4 10B Vol...

Page 28: ...nual for Handheld Pulse Oximeter 23 Notes When the Auto Power Off is set to On option if there is no Figure 4 10G TEMP unit setup Figure 4 10I Version info Figure 4 10H Default setting Figure 4 10J De...

Page 29: ...ve the cursor on Alert and press Confirm key then the oximeter enters into alert setting screen as shown in figure 4 11 Figure 4 11 Alert setting screen SpO2 Lo Limit SpO2 low limit setting range 50 9...

Page 30: ...then the over limit alert event will be activated that s the alert sound bibibibi occurs and the corresponding reading s blinks When measured on pediatric if the SpO2 reading is lower than or equal t...

Page 31: ...User Manual for Handheld Pulse Oximeter 26 Figure 4 12 Help information SpO2 measurement...

Page 32: ...Handheld Pulse Oximeter 27 Figure 4 12 Help information TEMP measurement 4 4 Record 4 4 1 Data Recall On main default screen short time press Record Back key to enter into data recall screen as shown...

Page 33: ...ecord Spot check Record list shows the recording time SpO2 value and pulse rate value as shown in figure 4 14 Figure 4 14 Spot check Record list If Continuos Record is selected then the screen shows t...

Page 34: ...d you need to review Select one record you need to review and press Confirm key then the screen shows the corresponding User User ID and trend graph as shown in figure 4 16 Figure 4 15 Continuos recor...

Page 35: ...st screen shown in figure 4 14 or 4 15 move the cursor on the record you want to delete and longtime pressing Mute Right key then an message Are you sure to delete all prompts on the screen as shown i...

Page 36: ...4 4 3 Data Upload If you want to upload the stored data SpO2 PR and TEMP values to the computer then Make sure the provided USB data cable is well connected between the device and PC before uploading...

Page 37: ...ngth LED sensor with wavelength Red light 663 nm Infrared light 890 nm Maximal average optical output power 2mW Display range 35 99 Measuring range 35 100 Measuring accuracy Arms is not greater than 3...

Page 38: ...6kPa Note portable and mobile RF communications equipment may affect the performance of the Oximeter I Low Perfusion Performance The accuracy of SpO2 and PR measurement still meet the precision descri...

Page 39: ...ctivated the value that is over limit will blink at the same time 6 2 Over limit indication sound mute setting During the measurement if the over limit indication sound is set to on short time press M...

Page 40: ...e See the package in your hand for detailed items and quantity 2 All the parts of the device should NOT be replaced at will If necessary please use the components provided by the manufacture or those...

Page 41: ...humidity control and 70 at a relative humidity up to 93 non condensing The oximeter is calibrated in the factory before sale there is no need to calibrate it during its life cycle However if it is ne...

Page 42: ...e into the device Device will not switch on 1 The batteries are drained or almost drained 2 The batteries are not inserted properly 3 The device is malfunctioning 1 Recharge or change batteries 2 Rein...

Page 43: ...o not shake the finger and keep calm during the measurement Please also avoid the oximeter and the cuff on the same limb for blood pressure and oxygen saturation measurement simultaneously 6 Q How to...

Page 44: ...are d Excessive patient movement e Placement of a sensor on an extremity with a blood pressure cuff arterial catheter or intravascular line f Exposure to the chamber with High pressure oxygen g There...

Page 45: ...he oxygen saturation PI Perfusion Index bpm Pulse rate Unit beats per minute Pulse bar graph Low battery voltage Battery is full Speaker mute icon Speaker volume icon SpO2 spot check record memory ful...

Page 46: ...re probe connector Power Left Key Mode Right Key Rotate lock Up Key Setting Down Key Menu Confirm key or Record Back key CE mark SN Serial number Date of manufacture Manufacturer including address Wit...

Page 47: ...directly proportional with its density or concentration When the light with certain wavelength emits on human tissue the measured intensity of light after absorption reflecting and attenuation in tiss...

Page 48: ...reater than 94 so the values below 94 are determined as hypoxia SpO2 90 is considered as the default threshold for determining anoxia by most researchers so SpO2 low limit of the oximeter is set as 90...

Page 49: ...thology reason Hypoxemia disease functional lack of HbO2 Pigmentation or abnormal oxyhemoglobin level Abnormal oxyhemoglobin variation Methemoglobin disease Sulfhemoglobinemia or arterial occlusion ex...

Page 50: ...User Manual for Handheld Pulse Oximeter 45 Quality Inspection Certificate...

Reviews: