CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-9
3C.3.1 Entering Into Programming Mode (Cont)
Use these steps to access the
GENERATOR SETUP
menu (membrane console).
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
Start with the generator switched OFF.
2.
While pressing and holding the
MENU
button, press the power
ON
button on the
console.
3.
Enter the password by pressing the
button sequence: [
1
] - [
8
] - [
4
] - [
5
].
4.
The
GENERATOR SETUP
menu will be
displayed next.
3C.4.0 GENERATOR SETUP MENU
The
GENERATOR SETUP
menu for the membrane console is shown below.
* GENERATOR SETUP *
UTILITY
APR EDITOR: DISABLED
GEN CONFIGURATION
DATA LINK
EXIT SETUP