CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-5
5.2.3 Error
Messages
These messages indicate that an error has occurred. The errors are logged in the error log; previous errors should be reviewed by service
personnel before taking further action.
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
APR MEMORY
ERROR
APR data has been corrupted.
See note 1 near end of this
section.
The license you entered
does not appear to be
valid. Please re-enter
your license key
The touchscreen software license
key has been entered incorrectly.
Verify and re-enter the license
key.
E003
GEN NVRAM ERR.
Generator CPU NVRAM
error
Generator CPU NVRAM data has
been corrupted.
Re-initialize generator CPU
NVRAM using generator
factory defaults.
E004
GEN RTC ERROR
Generator CPU Real
Time Clock error
Generator CPU real time clock is not
functioning.
Reset time and date.
E005
PS CONTACT ERR.
Main Contactor Error
Main contactor in generator did not
energize.
Call product support.
E006
ROTOR FAULT
Rotor Fault
1. Rotor starter may have detected
a current fault in the stator.
2. Generator was not ready to start
rotor.
Power unit off and retry rotor
start.
E007
FILAMENT FAULT
Filament Fault
Generator has detected filament
current <2 amps.
1. Check for open filament in
X-ray tube.
2. Check for poor connections
in the cathode cable.
3. Check fuses on filament
board(s).