CMP 200
®
Service Manual Ch # 901476-05 Rev. A
Page 3D-37
3D.9.0 PRECALIBRATION
SETUP
3D.9.1 AEC Setup Worksheet
Before continuing, it is suggested that a copy of the table below be filled in with all required information.
Refer to the example AEC setup worksheet on the next page.
FUNCTION
RECEPTOR 1
RECEPTOR 2
RECEPTOR 3
Film/Screen
1.
2.
3.
Nominal optical density:
Grid
ratio/SID:
Min - max kV range:
±
Density steps
Density mAs
(dose) change
+%:
(per step)
-%:
Chamber
type:
Regulatory AEC dose
requirements?
Is film processing maintained?
Assigned receptor name:
Are all cassettes similar?
Additional
notes:
Additional
notes:
Table 3D-11: AEC setup worksheet
ALL RECEPTORS MUST HAVE THE SAME NUMBER OF DENSITY STEPS AND THE SAME DENSITY DOSE
CHANGE PER STEP (DENSITY SETTINGS ARE COMMON TO ALL FILM SCREENS AND RECEPTORS).