3C.10.2 DAP SETUP Menu Items
Definitions of
DAP SETUP
menu items.
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
DAP DAP
checkbox
Enables or disables the DAP device.
Membrane console:
ON:
The DAP function is enabled.
OFF:
The DAP function is disabled.
In GenWare
®
, the DAP function is enabled or disabled via the
DAP
checkbox.
DEVICE TYPE 1
Device Type 1
Selects the DAP device.
0
= PTW PX-T11020.
1
= Gammex RMI 841S.
2
= VacuTec VacuDAP 2004.
3
= Scanditronix-Wellhoefer 120-131.
CAL. VALUE 1
Calibration Value
Allows the DAP reading to be calibrated by adjusting this parameter.
TEST VALUE 1
Test Value 1
A numeric value, supplied by the DAP device manufacturer, that represents the number of
pulses generated by the DAP device during TEST mode. Typical terms used by DAP device
manufacturers for this function are “Test Pulses”, “Test Value”, or “Test Count”, but other
names may be used.
The generator counts the number of pulses generated by the DAP device during TEST mode,
and reports a DAP failure error message if the actual number of test pulses are not the same as
the manufacturer-supplied test count, within an allowable margin of error.
DAP PRINTER
DAP Printer
Allows selection of the DAP printer type, or disabling of the DAP printer function. This is only
available with the DAP printer option.
OFF:
Disables the DAP printer function.
DPU414
:
Selects the Seiko Instruments DPU-414
printer.
SLP200:
Selects the Seiko Instruments SLP-200
printer.