Electromagnetic Compatibility (EMC)
Service Manual
13-5
13.7.3
Electromagnetic Immunity
Table 13-2.
Electromagnetic Immunity Guidelines
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
(IEC/EN 60601-1-2:2007, Table 2)
The warming system is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the warming system should assure that it is used in such an environment.
Immunity
Test
IEC/EN 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electric fast
transient/burst
IEC/EN 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for
power supply lines
± 1 kV for input/
output lines
Mains power quality should be
that of a typical commercial and/
or hospital environment.
Surge
IEC/EN 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial and/
or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
IEC/EN 61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
<5% UT
(>95% dip in UT)
for 0.5 cycle
Mains power quality should be
that of a typical commercial and/
or hospital environment.
If the user of the warming
system requires continued oper-
ation during power mains inter-
ruption, it is recommended that
the warming system be powered
from an uninterruptible power
supply or battery.
40% UT
(60% dip in UT) for
5 cycles
40% UT
(60% dip in UT) for
5 cycles
70% UT
(30% dip in UT) for
25 cycles
70% UT
(30% dip in UT) for
25 cycles
<5% UT
(>95% dip in UT)
for 5 seconds
<5% UT
(>95% dip in UT)
for 5 seconds
Power frequency
(50/60 Hz) magnetic
field
IEC/EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
Note:
UT is the AC main’s voltage prior to application of the test level.
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......