![Covidien Valleylab FT Series User Manual Download Page 125](http://html1.mh-extra.com/html/covidien/valleylab-ft-series/valleylab-ft-series_user-manual_2673869125.webp)
Valleylab FT10 Energy Platform User’s Guide
9-3
Cleaning
Maintenance and Repair
Cleaning
1. Turn off the system and unplug the power cord from the wall outlet.
2. Thoroughly wipe all surfaces of the energy platform and power cord with a damp cloth
and mild cleaning solution or disinfectant. The energy platform will withstand the
effects of cleaning over time without degrading the enclosure or display quality.
Product Service
Covidien recommends that all VLFT10GEN systems be returned to the manufacturer for all
service requirements. If any service is required without returning the system to the
manufacturer, Covidien recommends that only qualified personnel service the VLFT10GEN
system.
Covidien defines qualified personnel as a person with electrosurgical equipment repair
experience, such as biomedical personnel, and/or individuals who have taken official
Covidien training courses.
Warning
Shock Hazard
Before cleaning or servicing the unit, disconnect the power plug from the power
outlet in order to completely isolate the generator from mains power.
Notice
Do not clean the energy platform with abrasive cleaning or disinfectant compounds, solvents, or
other materials that could scratch the panels or damage the energy platform.
Do not spray cleaning fluids directly on the generator as damage to the generator may occur.