General information
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
1-17
Warning
Due to excessive restriction of the Air Liquide™*, Australian, and Dräger™* hose assemblies, reduced FIO
2
levels may result when oxygen inlet pressures
< 50 psi (345 kPa) are employed. Make sure oxygen inlet pressure
is ≥ 50 psi (345 kPa) when using these hose assemblies, to maintain correct FIO
2
levels.
NOTE:
To ensure that compliance compensation functions correctly, the
user must run EST or SST with the circuit configured as intended for
use on the patient.
Oxygen regulator pressure transducer
(P
O
) calibration tool
For use with:
•
PTS 2000: 4-079050-00
•
G-061541-SP
Calibrating oxygen regulator
pressure transducer
Cable ties, small
G-061096-00 or local supplier
Various places
Cutting tool
Local supplier
Various service procedures (for
cutting cable ties or wires)
Leak test fluid
4-004489-00
O
2
adapter assembly
Rivets
G-061182-00
Attaching serial number plate
Isopropyl alcohol
Local supplier
General cleaning
Cotton swabs
Local supplier
General cleaning
Touch-up paint
•
White liquid lacquer
•
Charcoal liquid lacquer
•
Gray liquid lacquer
•
G-061999-00
•
G-062000-00
•
G-061361-00
•
Ventilator cabinet
•
Ventilator lid
•
Ventilator cart
Electrostatic-shielding bags
•
66 x 46 cm (26 x 18 in.)
*
•
13 x 20 cm (5 x 8 in.)
*
•
28 x 38 cm (11 x 15 in.)
*
•
G-061534-00
•
G-061532-00
•
G-061533-00
•
UI display PCB
•
Optoswitch
•
Controller PCB, BBU PCB,
pressure solenoid PCB
Table 1-7: Tools, equipment, and service materials (continued)
Description
Manufacturer and model or
Covidien part number
Where used
Summary of Contents for Puritan Bennett 700 Series