General information
1-8
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
PEEP/CPAP pressure
(760 Ventilator only)
Sensing position: Exhalation limb
Range: 0 to 140 cmH
2
O (0 to 140 hPa)
Type: Silicon solid state
Rate
Sensing position: Microprocessor
Range: 1 to 500 breaths/min
Type: Calculated from inspiratory and expiratory time measurements
I:E ratio
Sensing position: Microprocessor
Range: 1:99.9 to 9.9:1
Type: Calculated from inspiratory and expiratory time measurements
Inspiratory time
(760 Ventilator only)
Sensing position: Microprocessor
Range: 0.1 to 9.9 s
Type: Inspiratory time measurement
Exhaled tidal volume
Sensing position: Exhalation limb
Range: 0 to 9 L
Type: Differential pressure pneumotach
Total minute volume
Sensing position: Exhalation limb
Range: 0 to 99 L
Type: Differential pressure pneumotach
Delivered volume
(760 Ventilator only)
Sensing position: Exhalation limb
Range: 0 to 3000 ml
Type: Differential pressure pneumotach
Spontaneous minute
volume
(760 Ventilator only)
Sensing position: Exhalation limb
Range: 0 to 99 L
Type: Calculated from expiratory flow measurements
FIO
2
sensor
Sensing position: Inspiration manifold
Range: 18 to 109 %
Type: Galvanic cell
Sensor life: 2 years, nominal. (Actual life depends on operating environment; operation at
higher temperature or FIO
2
levels will result in shorter sensor life.)
NOTE:
To achieve compliance with the more rigid requirements for oxygen monitoring (as
set forth in ISO 7767), Covidien recommends using an external monitor that meets
that standard.
Service specifications
Self-test capabilities
Automatic power-on self-test (POST): Verifies integrity of electronics.
Short self-test (SST): An abbreviated version of EST to be run by the operator. Characterizes
system leaks and system/tubing compliance.
Extended self-test (EST): Lets the technician thoroughly test the operational integrity of
the ventilator, both electronics and pneumatics. Tests can also be run individually, in
diagnostic EST mode.
Automatic, ongoing software and hardware checks.
Bacteria filter efficiency
Inspiratory and expiratory filters (disposable and reusable): 99.97% retention of 0.3
μ
m
particles at 100 L/min.
Languages
Software and user labeling available in Dutch, English, French, German, Italian, Japanese,
Polish, Portuguese, Russian, or Spanish.
Table 1-1: Specifications (continued)