Service and utility functions
4-10
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Settings for some serial devices include:
• Puritan Bennett™ CliniVision
HandHeld
and
VentNet
systems: 9600 Baud, 8 data bits, odd
parity.
• Hewlett-Packard Merlin patient monitor: 9600 Baud, 7 data bits, even parity.
• SpaceLabs™* patient monitor: 9600 Baud, 8 data bits, odd parity.
NOTE:
Settings and other specifications for external devices are subject to change. Consult the
external device manufacturer for the most current setting information.
4.2.3.2 Run tests
This function lets you run individual EST tests, perform ventilator calibrations, and other tests.
To access
Run tests
functions from the
Diagnostics/Calib
prompt, turn the knob until
Run
tests
is displayed, then press ACCEPT. Then turn the knob to locate the desired
Run tests
function.
4.2.3.2.1 EST tests: Running individual EST tests (diagnostic EST)
This function lets you run EST tests individually. It operates almost identically to EST, except
that:
• Test results are not written to NVRAM, and
• A test failure does not cause a VENT INOP condition to be declared. Thus, a ventilator that
fails an EST test can continue to operate
for troubleshooting purposes only
.
• Pressing ALARM RESET while running diagnostic EST returns you to the start of the full
(non-diagnostic) EST.
NOTE:
• Covidien recommends you always run the full EST before placing the ventilator into
operation following service. For preliminary troubleshooting purposes, however, you
may want to run EST tests individually. The
EST tests
function does not write error
information into NVRAM nor declare a VENT INOP condition when tests are failed.
• This
EST tests
function does
not
substitute for the full EST. After a ventilator is
serviced, it must be tested using the full EST before being returned to service.
Message window display
Description
EST tests
Press ACCEPT to proceed. Respond to the prompts, as indicated in Section 3.4.3,
until the name of the first EST test is displayed.
name of EST test
Run test ?
CLEAR (N) ACCEPT (Y)
Press ACCEPT to run this test, or press CLEAR to skip to the next test. If you choose
to run the test, the test’s pass/fail status is displayed at the end of the test.
EST finished testing
Press ACCEPT to proceed.
Unblock wye
Remove stopper from patient wye. Press ACCEPT. The ventilator now runs POST
and is ready for ventilation. To return to the service menu, hold down the MENU
key while POST is running.