4.2.3.1 Enable serial port
This function lets you enable the serial port to transfer data between the ventilator and a
computer. The serial port will remain enabled until the ventilator is powered off. If the
ventilator has a communications port, you can either enable the serial port or configure it.
Table 4-9: Performing calibrations/Updating NVRAM Constants
When this occurs
Do the following
Oxygen sensor replaced
Perform
FIO
2
sensor calib
function.
Oxygen regulator pressure transducer
replaced
Perform
O
2
pressure calib
function.
PEEP pump replaced
Perform
PEEP pump calib
function.
Oxygen solenoid assembly replaced
Perform
Update constants
function.
Flow sensor replaced
Perform
Update constants
function.
Piston/cylinder assembly replaced
Perform
Update constants
function.
Pressure solenoid PCB replaced
Perform
Update prsol nvram
,
FIO
2
sensor calib,
and
O
2
pressure calib
functions.
Caution
To ensure correct function of the pressure
solenoid PCB, be sure to perform the
Update
prsol nvram
function
before
the
FIO
2
sensor calib
and
O
2
pressure calib
functions.
P
e
P
cyl
compare test in EST fails
Perform
P
e
P
cyl
gain calib
function.
Message window display
Description
Serial ports
Press ACCEPT to proceed.
Configure port A
or
Configure port B
Press ACCEPT to proceed. Turn knob to change values of B (baud),
D (databits), or P (parity) and press ACCEPT to proceed. These settings
do not affect the serial port on the controller PCB, whose
configuration is set at 19200 Baud, 8 data bits, and no parity.
Enable up/download
Press ACCEPT to enable. Once up/downloading is enabled, you can use
one serial port to transfer service data:
•
serial port A only, if the Communications option is installed.
•
only the serial port on the controller PCB, if the Communications
option is
not
installed.
When up/downloading is enabled, normal operation of the
Communications panel ports is disabled. The up/downloading function
is disabled once ventilator settings are accepted or you turn the
ventilator power switch off and back on again.