Theory of operation
2-64
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Figure 2-53. Breath cycle for a VCV breath without PEEP
0 L/min
P
e
P
cyl
P
e
PEEP = 10 cmH
2
O
P
cyl
7-00165
Breath delivery
Exhalation piston
retraction
Exhalation valve closed
Breath
trigger
Piston
retraction
Leak compensation
Compliance compensation
Desired flow = 60 L/min
Exhalation
solenoid
Exhalation
flow
Piston
velocity