Specifications
A-2
Puritan Bennett 540™ Ventilator User’s Manual
A.3
Indicators and Alarms
Table A-3.
Internal Lithium Ion Battery
Voltage
25.2 VDC
Full-load capacity
4.8 Ah
Ampere-hour rating
On standby: 1.5 Ah
During ventilation: 0.5 Ah
Charging current
•
Standby mode
•
Ventilation mode
1.5 A/hr. (duration: < 6 hr.)
0.5 A/hr. (duration: < 13 hr.)
Average operating time at 25 °C (± 5 ºC)
with a fully charged battery (having less than
50 charge/discharge cycles) at the following displayed values:
Vt = 200 ml (± 5 ml), PIP = 10 cmH
2
O (± 2 cmH
2
O), Rtot = 20 bpm
Vt = 300 ml (± 5 ml), PIP = 20 cmH
2
O (± 2 cmH
2
O), Rtot = 15 bpm
Vt = 500 ml (± 5 ml), PIP = 30 cmH
2
O (± 2 cmH
2
O), Rtot = 15 bpm
Vt = 750 ml (± 5 ml), PIP = 45 cmH
2
O (± 2 cmH
2
O), Rtot = 20 bpm
(maximum settings)
11 hr. (–10%)
9 hr. (–10%)
6.5 hr. (–10%)
4.5 hr. (–10%)
Table A-4.
Remote Alarm
Remote Alarm
Port:
Also known as the Nurse’s Call port, it
provides for remote alerts of ventilator alarm
conditions.
An example of a setting that requires such a
feature is when the ventilator is used in an
isolation room.
The ventilator signals an alarm using a
normally open (NO) or a normally closed
(NC) signal.
A remote alarm is activated when an alarm
condition occurs, unless either of the
following is true:
•
Audio paused function is active
•
Ventilator power switch is turned off
The remote alarm port is an 8-pin female
connector. Allowable current is 100 mA at
24 VDC (maximum).
1
2
3
4
5
6
7
8
Nurse call pin-out (view from back of ventilator)
Pin
Signal
Wire color
1
relay common
black
2
normally closed (NC)
orange
3
normally open (NO)
brown
4
remote supply -(not
used)
-
5
RX Signal (not used)
-
6
TX Signal (not used)
-
7
remote (not
used)
-
8
not used
-
Summary of Contents for Puritan Bennett 540
Page 1: ...Puritan Bennett 540 Ventilator User s Manual...
Page 6: ...This page intentionally blank...
Page 8: ...This page intentionally blank...
Page 10: ...This page intentionally blank...
Page 94: ...This page intentionally blank...
Page 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Page 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Page 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Page 140: ......