![Covidien Newport HT70 Operator'S Manual Download Page 99](http://html1.mh-extra.com/html/covidien/newport-ht70/newport-ht70_operators-manual_2673909099.webp)
Battery Alarms
Operator's Manual
6-11
WARNING:
Immediately secure an external power source or insert a fully charged Power Pac battery pack
when the Backup Battery Shutdown Imminent alarm occurs.
Note:
It is highly recommended to carry at least one extra, fully charged Power Pac during transport or outdoor
applications.
6.6.7
Backup Battery Failure Alarm
The Backup Battery Failure alarm indicates that the backup battery is faulty and will not safely
operate the ventilator. Do not use the ventilator on battery power until it has been serviced.
6.6.8
Backup Battery Low Charge Alarm
The Backup Battery Low Charge alarm indicates that the backup battery has insufficient charge to
maintain ventilation if the Power Pac battery runs low or is removed. Attach to external power to
charge both batteries. If the backup battery does not charge within 3 hours, do not use the ven-
tilator on battery power until it has been serviced.
6.6.9
Power Pac Battery Pack Temperature Alarm
The Power Pac Battery Pack Temperature alarm indicates that the Power Pac battery pack tem-
perature has exceeded specifications for the battery. Replace the Power Pac battery pack with a
fully charged one. Contact Covidien Technical Services for repair or replacement of the battery.
6.6.10
Backup Battery Temperature Alarm
The Backup Battery Temperature alarm message indicates that the backup battery temperature
has exceeded specifications for the battery. Connect the ventilator to external power; do not use
on battery power until the ventilator has been serviced. Use an alternate means of ventilation and
have the ventilator serviced.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......