Ventilator Alarms
6-8
Operator's Manual
6.5.3
Occlusion Alarm
An Occlusion alarm is activated by an obstruction in the breathing circuit. The ventilator will
attempt to relieve the pressure that has built up in the circuit and will not deliver additional
breaths until the situation is resolved. The alarm resets when the occlusion is resolved and breath
delivery will resume at that point.
6.5.4
Sustained Occlusion Alarm
A Sustained Occlusion alarm is activated if the Occlusion alarm is not cleared within 10 seconds
or two breath periods, whichever is shorter. The ventilator will attempt to relieve the pressure and
will not deliver additional breaths until the situation is resolved. The alarm resets when the occlu-
sion is resolved and breath delivery will resume at that point.
6.5.5
Check Circuit or Prox Line Alarm
The Circuit Check or Prox Line alarm indicates that the circuit has become disconnected or the
proximal pressure tubing is disconnected, kinked, or has water in it. Check the circuit for discon-
nects or problems with the prox line pressure tubing/prox line filter.
Note:
Be sure to keep the prox inline filter clean and dry at all times.
6.5.6
Flow Sensor Disconnect/Failure Alarms
The Flow Sensor Disconnect and Failure alarms indicate that the on-airway flow sensor has
become disconnected, the tubing has been partially blocked with water, or the flow sensor is no
longer w
+ only available on the HT70PM when the on-airway flow sensor is in use
6.5.7
Pressure Control Setting Not Reached Alarm
The Pressure Control Setting Not Reached alarm is activated by inadequate pressure rise during
pressure-controlled breaths. Check for leaks and that the slope/rise is set to a fast enough level.
6.5.8
No External Power Alarm
The No External Power alarm is activated by disconnection from the power cord or a power inter-
ruption. The ventilator will automatically switch to the Power Pac battery pack or backup battery.
Pressing the
Audio Paused/Reset
button will clear the alarm.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......